The accuracy evaluation of four blood glucose monitoring systems according to ISO 15197:2003 and ISO 15197:2013 criteria.

Primary Care Diabetes 2019 Februrary 13
Self-monitoring of blood glucose is recommended to monitor glycaemic levels for the prevention and treatment of diabetes mellitus.The accuracy of blood glucose monitoring systems (BGMS) directly affects the health of the patients. In this study, we evaluated the accuracy of four different blood glucose monitoring systems, including Gold-Accu Meter, Nova StatStrip Xpress® Glucose Hospital Meter, Freestyle Optium Neo H® and Accu-Chek Performa according to ISO15197:2003 and ISO15197:2013 criteria. We found that four brands of BGMS (eight BGMS) all met the minimum requirements for system accuracy performance of ISO 15197:2003 and the accuracy criteria of ISO 15197:2013. All results of BGMS lie within zones A and B of the Consensus Error Grid (CEG). In conclusion, all BGMS met the accuracy criteria of ISO 15197:2003 and the accuracy criteria of ISO 15197:2013.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app