Technical Note: A comparison of point set registration methods for electromagnetic tracking.
Medical Physics 2019 May
PURPOSE: High dose rate brachytherapy applies intense and destructive radiation. A treatment plan defines radiation source dwell positions to avoid irradiating healthy tissue. The study discusses methods to quantify any positional changes of source locations along the various treatment sessions.
METHODS: Electromagnetic tracking (EMT) localizes the radiation source during the treatment sessions. But in each session the relative position of the patient relative to the filed generator is changed. Hence, the measured dwell point sets need to be registered onto each other to render them comparable. Two point set registration techniques are compared: a probabilistic method called coherent point drift (CPD) and a multidimensional scaling (MDS) technique.
RESULTS: Both enable using EMT without external registration and achieve very similar results with respect to dwell position determination of the radiation source. Still MDS achieves smaller grand average deviations (CPD-rPSR: MD = 2.55 mm, MDS-PSR: MD = 2.15 mm) between subsequent dwell position determinations, which also show less variance (CPD-rPSR: IQR = 4 mm, MDS-PSR: IQR = 3 mm). Furthermore, MDS is not based on approximations and does not need an iterative procedure to track sensor positions inside the implanted catheters.
CONCLUSION: Although both methods achieve similar results, MDS is to be preferred over rigid CPD while nonrigid CPD is unsuitable as it does not preserve topology.
METHODS: Electromagnetic tracking (EMT) localizes the radiation source during the treatment sessions. But in each session the relative position of the patient relative to the filed generator is changed. Hence, the measured dwell point sets need to be registered onto each other to render them comparable. Two point set registration techniques are compared: a probabilistic method called coherent point drift (CPD) and a multidimensional scaling (MDS) technique.
RESULTS: Both enable using EMT without external registration and achieve very similar results with respect to dwell position determination of the radiation source. Still MDS achieves smaller grand average deviations (CPD-rPSR: MD = 2.55 mm, MDS-PSR: MD = 2.15 mm) between subsequent dwell position determinations, which also show less variance (CPD-rPSR: IQR = 4 mm, MDS-PSR: IQR = 3 mm). Furthermore, MDS is not based on approximations and does not need an iterative procedure to track sensor positions inside the implanted catheters.
CONCLUSION: Although both methods achieve similar results, MDS is to be preferred over rigid CPD while nonrigid CPD is unsuitable as it does not preserve topology.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app