Validation of Post-Treatment PET-Based Dosimetry Software for Hepatic Radioembolization of Yttrium-90 Microspheres.

Medical Physics 2019 Februrary 12
PURPOSE: Yttrium-90 (90 Y) microsphere radioembolization enables selective internal radiotherapy for hepatic malignancies. Currently, there is no standard post-delivery imaging and dosimetry of the microsphere distribution to verify treatment. Recent studies have reported utilizing the small positron yield of 90 Y (32 ppm) with PET to perform treatment verification and dosimetry analysis. In this study, we validated a commercial dosimetry software, MIM SurePlan™ LiverY90 (MIM Software Inc., Cleveland, OH), for clinical use.

METHODS: A MATLAB-based algorithm for 90 Y PET-based dosimetry was developed in-house and validated for the purpose of commissioning the commercial software. The algorithm is based on voxel S values and dosimetry formalism reported in MIRD Pamphlet 17. We validated the in-house algorithm to establish it as the ground truth by comparing results from a digital point phantom and a digital uniform cylinder to manual calculations. Once we validated our in-house MATLAB-based algorithm, we used it to perform acceptance testing and commissioning of the commercial dosimetry software, MIM SurePlan, which uses the same dosimetry formalism. A 0.4cm/5% gamma test was performed on PET-derived dose maps from each algorithm of uniform digital and non-uniform physical phantoms filled with 90 Y chloride solution. Average dose (Davg ) and minimum dose to 70% (D70) of a given volume of interest (VOI) were compared for the digital phantom, the physical phantom, and 5 patient cases (27 tumor VOIs), representing different clinical scenarios.

RESULTS: The gamma-pass rates were 97.26% and 97.66% for the digital and physical phantoms, respectively. The differences between Davg and D70 were 0.076% and 0.10% for the digital phantom respectively, and <5.2% for various VOIs in the physical phantom. In the clinical cases, 96.3% of the VOIs had a difference <5% for Davg , and 88.9% of the VOIs had a difference <5% for D70.

CONCLUSIONS: Dose calculation results from MIM SurePlan were found to be in good agreement with our in-house algorithm. This indicates that MIM SurePlan performs as it should and, hence, can be deemed accepted and commissioned for clinical use for post-implant PET-based dosimetry of 90 Y radioembolization. This article is protected by copyright. All rights reserved.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app