Cardiac output measurement during exercise in COPD: A comparison of dye dilution and impedance cardiography.
Clinical Respiratory Journal 2019 April
INTRODUCTION: Impedance cardiography (IC) derived from morphological analysis of the thoracic impedance signal is now commonly used for noninvasive assessment of cardiac output (CO) at rest and during exercise. However, in Chronic Obstructive Pulmonary Disease (COPD), conflicting findings put its accuracy into question.
OBJECTIVES: We therefore compared concurrent CO measurements captured by IC (PhysioFlow: COIC ) and by the indocyanine green dye dilution method (CODD ) in patients with COPD.
METHODS: Fifty paired CO measurements were concurrently obtained using the two methods from 10 patients (FEV1 : 50.5 ± 17.5% predicted) at rest and during cycling at 25%, 50%, 75% and 100% peak work rate.
RESULTS: From rest to peak exercise COIC and CODD were strongly correlated (r = 0.986, P < 0.001). The mean absolute and percentage differences between COIC and CODD were 1.08 L/min (limits of agreement (LoA): 0.05-2.11 L/min) and 18 ± 2%, respectively, with IC yielding systematically higher values. Bland-Altman analysis indicated that during exercise only 7 of the 50 paired measurements differed by more than 20%. When data were expressed as changes from rest, correlations and agreement between the two methods remained strong over the entire exercise range (r = 0.974, P < 0.001, with no significant difference: 0.19 L/min; LoA: -0.76 to 1.15 L/min). Oxygen uptake (VO2 ) and CODD were linearly related: r = 0.893 (P < 0.001), CODD = 5.94 × VO2 + 2.27 L/min. Similar results were obtained for VO2 and COIC (r = 0.885, P < 0.001, COIC = 6.00 × VO2 + 3.30 L/min).
CONCLUSIONS: These findings suggest that IC provides an acceptable CO measurement from rest to peak cycling exercise in patients with COPD.
OBJECTIVES: We therefore compared concurrent CO measurements captured by IC (PhysioFlow: COIC ) and by the indocyanine green dye dilution method (CODD ) in patients with COPD.
METHODS: Fifty paired CO measurements were concurrently obtained using the two methods from 10 patients (FEV1 : 50.5 ± 17.5% predicted) at rest and during cycling at 25%, 50%, 75% and 100% peak work rate.
RESULTS: From rest to peak exercise COIC and CODD were strongly correlated (r = 0.986, P < 0.001). The mean absolute and percentage differences between COIC and CODD were 1.08 L/min (limits of agreement (LoA): 0.05-2.11 L/min) and 18 ± 2%, respectively, with IC yielding systematically higher values. Bland-Altman analysis indicated that during exercise only 7 of the 50 paired measurements differed by more than 20%. When data were expressed as changes from rest, correlations and agreement between the two methods remained strong over the entire exercise range (r = 0.974, P < 0.001, with no significant difference: 0.19 L/min; LoA: -0.76 to 1.15 L/min). Oxygen uptake (VO2 ) and CODD were linearly related: r = 0.893 (P < 0.001), CODD = 5.94 × VO2 + 2.27 L/min. Similar results were obtained for VO2 and COIC (r = 0.885, P < 0.001, COIC = 6.00 × VO2 + 3.30 L/min).
CONCLUSIONS: These findings suggest that IC provides an acceptable CO measurement from rest to peak cycling exercise in patients with COPD.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app