A thin-film multichannel electrode for muscle recording and stimulation in neuroprosthetics applications.

OBJECTIVE: We propose, design and test a novel thin-film multichannel electrode that can be used for both recording from and stimulating a muscle in acute implants.

APPROACH: The system is built on a substrate of polyimide and contains 12 recording and three stimulation sites made of platinum. The structure is 420 µm wide, 20 µm thick and embeds the recording and stimulation contacts on the two sides of the polyimide over an approximate length of 2 cm. We show representative applications in healthy individuals as well as tremor patients. The designed system was tested by a psychometric characterization of the stimulation contacts in six tremor patients and three healthy individuals determining the perception threshold and current limit as well as the success rate in discriminating elicited sensations (electrotactile feedback). Also, we investigated the possibility of using the intramuscular electrode for reducing tremor in one patient by electrical stimulation delivered with timing based on the electromyographic activity recorded with the same electrode.

MAIN RESULTS: In the tremor patients, the current corresponding to the perception threshold and the current limit were 0.7  ±  0.2 and 1.4  ±  0.7 mA for the wrist flexor muscles and 0.4  ±  0.2 and 1.5  ±  0.7 mA for the extensors. In one patient, closed-loop stimulation resulted in a decrease of the tremor power  >50%. In healthy individuals the perception threshold and current limits were 0.9  ±  0.6 and 2.1  ±  0.6 mA for the extensor carpi radialis muscle. The subjects could distinguish four or six stimulation patterns (two or three stimulation sites  ×  two stimulation current amplitudes) with true positive rate  >80% (two subjects) and  >60% (one subject), respectively.

SIGNIFICANCE: The proposed electrode provides a compact multichannel interface for recording electromyogram and delivering electrical stimulation in applications such as neuroprostheses for tremor suppression and closed-loop myoelectric prostheses.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app