A randomized, controlled, multi-center trial of the efficacy and safety of the Occlutech Figulla Flex-II Occluder compared to the Amplatzer Septal Occluder for transcatheter closure of secundum atrial septal defects.
Catheterization and Cardiovascular Interventions 2019 Februrary 2
AIMS: The aim of this study was to compare the efficacy and safety of the Occlutech Figulla Flex II Occluder (OFFII) with the Amplatzer Septal Occluder (ASO) in patients > 8kg undergoing transcatheter ASD closure.
METHODS AND RESULTS: Randomized, controlled, multi-center prospective clinical trial with randomization 2:1 in favor of the OFFII. Primary efficacy endpoint was the rate of successful device placement and defect closure without major complications at hospital discharge. All data were assessed through a core laboratory. Interim analysis was performed when 70% of the patients were treated to evaluate for noninferiority. From a total of 176 randomized subjects, interim analysis was performed on the first 158 patients (65.2% female) (107 OFFII/51 ASO) undergoing device closure at a median weight of 42 kg (range 13-125 kg). Seventy-six percent (120 patients) completed 6-month follow-up. Successful device placement (first attempt) was achieved in 99.1% of the OFF group vs 90.2% of the ASO group (P < 0.05). Early efficacy success was achieved in 94.4% of the OFFII group vs 90.2% of the ASO group (P < 0.001). The incidence of major complications was 5.6% for the OFFII group compared to 9.8% for the ASO.
CONCLUSIONS: The OFFII device was not inferior to the ASO with less complications and greater efficacy than the ASO.
METHODS AND RESULTS: Randomized, controlled, multi-center prospective clinical trial with randomization 2:1 in favor of the OFFII. Primary efficacy endpoint was the rate of successful device placement and defect closure without major complications at hospital discharge. All data were assessed through a core laboratory. Interim analysis was performed when 70% of the patients were treated to evaluate for noninferiority. From a total of 176 randomized subjects, interim analysis was performed on the first 158 patients (65.2% female) (107 OFFII/51 ASO) undergoing device closure at a median weight of 42 kg (range 13-125 kg). Seventy-six percent (120 patients) completed 6-month follow-up. Successful device placement (first attempt) was achieved in 99.1% of the OFF group vs 90.2% of the ASO group (P < 0.05). Early efficacy success was achieved in 94.4% of the OFFII group vs 90.2% of the ASO group (P < 0.001). The incidence of major complications was 5.6% for the OFFII group compared to 9.8% for the ASO.
CONCLUSIONS: The OFFII device was not inferior to the ASO with less complications and greater efficacy than the ASO.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app