OBJECTIVES: To investigate biotin interference on three cardiac troponin (cTn) assays and demonstrate a method to overcome biotin interference.
METHODS: cTn levels were measured in (1) plasma from healthy volunteers on 10-mg daily biotin supplementation mixed with a plasma with known elevated troponin, (2) plasmas with known elevated cTn after mixing in reagent biotin to simulate supplementation, and (3) biotin-spiked plasma specimens pretreated with streptavidin-agarose beads.
RESULTS: Daily biotin ingestion (10 mg) and studies simulating daily biotin use resulted in significant interference in the Gen5 cardiac troponin T (cTnT) assay; the contemporary Gen 4 cTnT and high-sensitivity cardiac troponin I (hs-cTnI) assays were unaffected. The biotin interference threshold was 31, 315, and more than 2,000 ng/mL for Gen5 cTnT, cTnT, and hs-cTnI assays, respectively. Streptavidin pretreatment blocked biotin interference in cTn assays.
CONCLUSIONS: Biotin interference is possible at plasma concentrations achievable by ingestion of over-the-counter supplements that may lead to delayed or missed diagnosis of myocardial injury with the Gen5 cTnT assay.
METHODS: cTn levels were measured in (1) plasma from healthy volunteers on 10-mg daily biotin supplementation mixed with a plasma with known elevated troponin, (2) plasmas with known elevated cTn after mixing in reagent biotin to simulate supplementation, and (3) biotin-spiked plasma specimens pretreated with streptavidin-agarose beads.
RESULTS: Daily biotin ingestion (10 mg) and studies simulating daily biotin use resulted in significant interference in the Gen5 cardiac troponin T (cTnT) assay; the contemporary Gen 4 cTnT and high-sensitivity cardiac troponin I (hs-cTnI) assays were unaffected. The biotin interference threshold was 31, 315, and more than 2,000 ng/mL for Gen5 cTnT, cTnT, and hs-cTnI assays, respectively. Streptavidin pretreatment blocked biotin interference in cTn assays.
CONCLUSIONS: Biotin interference is possible at plasma concentrations achievable by ingestion of over-the-counter supplements that may lead to delayed or missed diagnosis of myocardial injury with the Gen5 cTnT assay.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app