Reliability of, and Agreement Between, two Breath-by-Breath Indirect Calorimeters at Varying Levels of Inspiratory Oxygen.
Nutrition in Clinical Practice 2019 Februrary 5
BACKGROUND: Indirect calorimetry (IC) is considered the accurate way of measuring energy expenditure (EE). IC devices often apply the Haldane transformation, introducing errors at inspiratory oxygen fraction (FiO2 ) >60%. The aim was to assess measurement reliability and agreement between an unevaluated IC (device 2) (Beacon Caresystem, Mermaid Care A/S, Noerresundby, Denmark) not using Haldane transformation and an IC that does (device 1) (Ecovx, GE, Helsinki, Finland) at varying FiO2 .
METHODS: Twenty healthy male subjects participated, with 16 completing the study (33 ± 9 years, 83.3 ± 16 kg, 1.83 ± 0.08 m). Subjects were mechanically ventilated in pressure support (3cmH2 O; positive end-expiratory pressure: 3cmH2 O) at FiO2 of 21%, 50%, 85%, and 21% for 15 minutes at each FiO2 . Mean EE, oxygen consumption (VO2 ), and CO2 production (VCO2 ) were compared within and between devices across FiO2 levels.
RESULTS: Device 2 showed within-device EE significant differences at 21% vs 50% FiO2 and device 1 for VCO2 at 50% vs. 85% FiO2 . For all variables, both devices showed reliable measurements at 21% and 50% FiO2 , but at 85%, FiO2 bias and limits of agreement increased. Between devices, there were significant differences for EE at both 21% and 85% FiO2 for VO2 and for VCO2 at 85% FiO2 .
CONCLUSION: Both systems measured EE, VO2 , and VCO2 at 21%-85% FiO2 reliably but with bias at 85% FiO2 . The devices were in agreement at 21% and 50% FiO2 , but further studies need to confirm accuracy at high FiO2 .
METHODS: Twenty healthy male subjects participated, with 16 completing the study (33 ± 9 years, 83.3 ± 16 kg, 1.83 ± 0.08 m). Subjects were mechanically ventilated in pressure support (3cmH2 O; positive end-expiratory pressure: 3cmH2 O) at FiO2 of 21%, 50%, 85%, and 21% for 15 minutes at each FiO2 . Mean EE, oxygen consumption (VO2 ), and CO2 production (VCO2 ) were compared within and between devices across FiO2 levels.
RESULTS: Device 2 showed within-device EE significant differences at 21% vs 50% FiO2 and device 1 for VCO2 at 50% vs. 85% FiO2 . For all variables, both devices showed reliable measurements at 21% and 50% FiO2 , but at 85%, FiO2 bias and limits of agreement increased. Between devices, there were significant differences for EE at both 21% and 85% FiO2 for VO2 and for VCO2 at 85% FiO2 .
CONCLUSION: Both systems measured EE, VO2 , and VCO2 at 21%-85% FiO2 reliably but with bias at 85% FiO2 . The devices were in agreement at 21% and 50% FiO2 , but further studies need to confirm accuracy at high FiO2 .
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app