Effect of pulse-width-modulated LED light on the temperature change of composite in tooth cavities.

Dental Materials 2019 April
OBJECTIVE: The purpose of this study was to investigate the effects of the radiant emittance and cure time of pulse width modulation (PWM)-controlled LED light on the temperature change of composite and dentin.

METHODS: Class I cavities (M-D 6mm, B-L 4mm, Depth 2.5mm) were prepared on 30 extracted human molars and vertically sectioned to expose the mesial side of the cavity and tooth. Cavities were filled with Bulk Fill Posterior Restorative (BFP, 3M ESPE) and cured with an LED light. The duty ratio (% of time the light is on) and cure time of the LED light were controlled using an Arduino UNO microcontroller (PWM) as follows (6 groups, n=5): 10%/100 s, 30%/33.3s, 50%/20s, 100%/10s, Increase mode (0→100%)/20s, and Decrease mode (100→0%)/20s. All measurements were performed at 100Hz PWM with the constant total radiant exposure. Thermograms of the specimens were recorded using an infrared thermal camera (VarioCamhr head 700, InfraTec GmbH) for a pre-cure time of 20s, cure time, and a post-cure time of 100s at room temperature of 30±0.5°C. Temperature change data on the composite and dentin surfaces were collected at incremental distances of 0.625mm and 1mm from the top of the cavity to the pulp. Data were statistically analyzed using two-way ANOVA and Tukey's post-hoc test at α=0.05.

RESULTS: A rapid temperature increase occurred within the cavity during light curing. The maximum temperature rises (ΔTmax ) were observed at 0.625mm apical from the top and middle of the cavity. The ΔTmax ranged from 7.62 to 16.74°C at 0.625mm apical from the top, 4.83 to 11.39°C at the floor of the cavity, and 3.16 to 8.09°C in the dentin 1mm beneath the cavity base. The ΔTmax of composite and dentin increased and the time to reach ΔT=5°C decreased with increasing duty ratio at constant radiant exposure. In the Increase mode, ΔTmax was lower than that of 50%/20s mode. The ΔTmax in the Decrease mode was similar to that of 100%/10s mode.

SIGNIFICANCE: The PWM-LED curing light system controlled by a microcontroller provided a useful tool of varying the radiant emittance and cure time with constant radiant exposure to evaluate temperature change of composite and dentin. These result will be helpful to determine proper curing modes with varying radiant emittance of the LED curing light for decreasing temperature change of composite and dentin. At constant radiant exposure and cure times, the Increase mode showed lower and slower temperature rises than the 50%/20s and Decrease mode. Within the limitations of this in vitro study, when radiant exposure is constant, a curing light with lower radiant emittance can induce relatively low thermal transfer, thereby decreasing the risk of pulpal damage.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app