Mechanical Properties of Various Glide Path Preparation Nickel-titanium Rotary Instruments.
Journal of Endodontics 2019 Februrary
INTRODUCTION: The aim of this study was to compare the cyclic fatigue, torsional resistance, and bending stiffness of single-file glide path preparation nickel-titanium instruments.
METHODS: ProGlider (#16/progressive taper; Dentsply Sirona, Ballaigues, Switzerland), One G (#14/.03; Micro-Mega, Besancon, France), and EdgeGlidePath (#16/progressive taper; Edge Endo, Albuquerque, NM) were tested. For the bending stiffness test, the instrument was bent to a 45° angle along the main axis while 3 mm of the tip was secured. Cyclic fatigue resistance was measured during pecking movement, whereas the file was rotated within a reproducible simulated canal with a 3-mm radius and a 90° angle of curvature. The ultimate torsional strength and toughness were evaluated using a custom device. While tightly holding the end of the file at 3 mm, a rotational load of 2 rpm speed was applied until fracture. The results were analyzed using 1-way analysis of variance and Tukey post hoc comparison. A microscopic inspection was performed on the surface of a separate instrument using a scanning electron microscope.
RESULTS: EdgeGlidePath showed superior cyclic fatigue resistance compared with the other systems. One G showed higher maximum torsional strength than the others. One G showed the largest distortion angle and the highest toughness followed by EdgeGlidePath and ProGlider (P < .05). One G also showed larger bending stiffness than the others, whereas EdgeGlidePath showed a larger residual angle than the others (P < .05).
CONCLUSIONS: Conventional wire showed higher toughness and torsional resistance than M-Wire and heat-treated nickel-titanium instruments. However, heat-treated wire showed higher cyclic resistance than conventional wire and M-Wire.
METHODS: ProGlider (#16/progressive taper; Dentsply Sirona, Ballaigues, Switzerland), One G (#14/.03; Micro-Mega, Besancon, France), and EdgeGlidePath (#16/progressive taper; Edge Endo, Albuquerque, NM) were tested. For the bending stiffness test, the instrument was bent to a 45° angle along the main axis while 3 mm of the tip was secured. Cyclic fatigue resistance was measured during pecking movement, whereas the file was rotated within a reproducible simulated canal with a 3-mm radius and a 90° angle of curvature. The ultimate torsional strength and toughness were evaluated using a custom device. While tightly holding the end of the file at 3 mm, a rotational load of 2 rpm speed was applied until fracture. The results were analyzed using 1-way analysis of variance and Tukey post hoc comparison. A microscopic inspection was performed on the surface of a separate instrument using a scanning electron microscope.
RESULTS: EdgeGlidePath showed superior cyclic fatigue resistance compared with the other systems. One G showed higher maximum torsional strength than the others. One G showed the largest distortion angle and the highest toughness followed by EdgeGlidePath and ProGlider (P < .05). One G also showed larger bending stiffness than the others, whereas EdgeGlidePath showed a larger residual angle than the others (P < .05).
CONCLUSIONS: Conventional wire showed higher toughness and torsional resistance than M-Wire and heat-treated nickel-titanium instruments. However, heat-treated wire showed higher cyclic resistance than conventional wire and M-Wire.
Full text links
Related Resources
Trending Papers
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Anti-Arrhythmic Effects of Heart Failure Guideline-Directed Medical Therapy and Their Role in the Prevention of Sudden Cardiac Death: From Beta-Blockers to Sodium-Glucose Cotransporter 2 Inhibitors and Beyond.Journal of Clinical Medicine 2024 Februrary 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app