Performance evaluation of a new and improved cuvette-based automated urinalysis analyzer with phase contrast microscopy.
BACKGROUND: The use of phase contrast in urinalysis has been highly recommended. A new system, sediMAX conTRUST PRO, is now available providing simultaneous automated phase contrast and bright field microscopy. This study aimed to evaluate both analytical and diagnostic performance of this new analyzer.
METHODS: Results from 504 samples evaluated with the sediMAX conTRUST PRO were compared to those obtained from the same samples by manual microscopy (MM). Analytical and diagnostic performance were assessed according to established guidelines.
RESULTS: The concentration of red blood cells (RBCs)and white blood cells (WBCs) at which the LoQ satisfied a CV < 25% was 12 particles per μL (p/μL) and 8 p/μL, respectively. Within one grade of agreement concordance was quite high, 97.8% for RBCs and 98.0% for WBCs, and above 90% for all other particles. Overall, diagnostic sensitivity and specificity were good (>80%) for the particles considered, although lower sensitivities, 70.6% and 61.8%, were respectively found for hyaline and pathological casts.
CONCLUSIONS: The sediMAX conTRUST PRO provides very good performance in terms of RBC and WBC recognition and enumeration, and quite good performance for all other particles. Hyaline cast and pathological cast identification is fine and comparable to other automated systems, but could use further improvement.
METHODS: Results from 504 samples evaluated with the sediMAX conTRUST PRO were compared to those obtained from the same samples by manual microscopy (MM). Analytical and diagnostic performance were assessed according to established guidelines.
RESULTS: The concentration of red blood cells (RBCs)and white blood cells (WBCs) at which the LoQ satisfied a CV < 25% was 12 particles per μL (p/μL) and 8 p/μL, respectively. Within one grade of agreement concordance was quite high, 97.8% for RBCs and 98.0% for WBCs, and above 90% for all other particles. Overall, diagnostic sensitivity and specificity were good (>80%) for the particles considered, although lower sensitivities, 70.6% and 61.8%, were respectively found for hyaline and pathological casts.
CONCLUSIONS: The sediMAX conTRUST PRO provides very good performance in terms of RBC and WBC recognition and enumeration, and quite good performance for all other particles. Hyaline cast and pathological cast identification is fine and comparable to other automated systems, but could use further improvement.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app