We have located links that may give you full text access.
Importance of monitoring zones in the detection of arrhythmias in patients with implantable cardioverter-defibrillators under remote monitoring.
Portuguese Journal of Cardiology : An Official Journal of the Portuguese Society of Cardiology 2019 January 25
INTRODUCTION: Implantable cardioverter-defibrillator (ICD) monitoring zones (MZ) provide passive features that do not interfere with the functioning of active treatment zones. However, it is not known for certain whether programming an MZ affects arrhythmia detection by the ICD. The aim of the present study is to assess the clinical relevance of MZ in a population of patients with ICDs.
METHODS: In this retrospective analysis of patients with ICDs, with or without cardiac resynchronization therapy, for primary prevention under remote monitoring, the MZ was analyzed and recorded arrhythmias were assessed in detail.
RESULTS: A total of 221 patients were studied (77% men; age 64±12 years). Mean ejection fraction was 30±12%. The mean follow-up was 63±35 months. One hundred and seventy-four MZ events were documented in 139 patients (62.9%): 74 of non-sustained ventricular tachycardia (NSVT), 42 of supraventricular tachycardia, 44 of atrial fibrillation/atrial flutter, and five cases of noise. Among the 137 patients who presented with arrhythmias in the MZ (excluding two cases with noise detection only), 22 (16.1%) received appropriate shocks and/or antitachycardia pacing (ATP), while of the other 84 patients, 15.5% received appropriate ICD treatment (p=NS). In patients who presented with NSVT in the MZ, 15 (20.5%) received appropriate shocks and/or ATP. In accordance with the MZ findings, physicians decided to change outpatient medication in 41.7% of all patients in whom arrhythmic events were reported.
CONCLUSION: Ventricular and supraventricular arrhythmias are common findings in the MZ of ICD patients. Programming an MZ is valuable in the diagnosis of arrhythmias and may be a useful tool in clinical practice.
METHODS: In this retrospective analysis of patients with ICDs, with or without cardiac resynchronization therapy, for primary prevention under remote monitoring, the MZ was analyzed and recorded arrhythmias were assessed in detail.
RESULTS: A total of 221 patients were studied (77% men; age 64±12 years). Mean ejection fraction was 30±12%. The mean follow-up was 63±35 months. One hundred and seventy-four MZ events were documented in 139 patients (62.9%): 74 of non-sustained ventricular tachycardia (NSVT), 42 of supraventricular tachycardia, 44 of atrial fibrillation/atrial flutter, and five cases of noise. Among the 137 patients who presented with arrhythmias in the MZ (excluding two cases with noise detection only), 22 (16.1%) received appropriate shocks and/or antitachycardia pacing (ATP), while of the other 84 patients, 15.5% received appropriate ICD treatment (p=NS). In patients who presented with NSVT in the MZ, 15 (20.5%) received appropriate shocks and/or ATP. In accordance with the MZ findings, physicians decided to change outpatient medication in 41.7% of all patients in whom arrhythmic events were reported.
CONCLUSION: Ventricular and supraventricular arrhythmias are common findings in the MZ of ICD patients. Programming an MZ is valuable in the diagnosis of arrhythmias and may be a useful tool in clinical practice.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app