A linear relationship for the LET-dependence of Gafchromic EBT3 film in spot-scanning proton therapy.

Radiochromic film (RCF) is a valuable dosimetric tool, primarily due to its sub-millimeter spatial resolution. For accurate proton dosimetry, the dependence of film response on linear energy transfer (LET) must be characterized and calibrated. In this work, we characterized film under-response, or 'quenching', as a function of dose-weighted linear energy transfer (LETd ) in several proton fields and established a simple, linear relationship with LETd . We performed measurements as a function of depth in a PMMA phantom irradiated by a spot-scanning proton beam. The fields had energies of 71.3 MeV, 71.3 MeV with filter, and 159.9 MeV. At each depth (measurements taken in depth step sizes of 0.5-1 mm in the Bragg peak), we measured dose with a PTW Markus ion chamber and EBT3 RCF. EBT3 under-response was characterized by the ratio of dose measured with film to that with ion chamber. LETd values for our experimental setup were calculated using in-house clinical Monte Carlo code. Measured film under-response increased with LETd , from approximately 10% under-response for LETd   =  5 keV µm-1 to approximately 20% for LETd   =  8 keV µm-1 . The under-response for all measurements was plotted versus LETd . A linear fit to the data was performed, yielding a function for under-response, [Formula: see text], with respect to LETd : [Formula: see text]. Finally, the linear under-response relationship was applied to a film measurement within a spread-out Bragg peak (SOBP). Without correcting for LETd -dependence in the SOBP measurement, the discrepancy between film and Monte Carlo profiles was greater than 15% at the distal edge. With correction, the corrected film profile was within 2% and 1 mm of the Monte Carlo profile. RCF response depends on LETd , potentially under-responding by  >15% in clinically-relevant scenarios. Therefore, it is insufficient to perform only a dose calibration; LET calibration is also necessary for accurate proton film dosimetry. The LETd -dependence of EBT3 can be described by a single, linear function over a range of clinically-relevant proton therapy LETd values.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app