Monte Carlo simulations of out-of-field skin dose due to spiralling contaminant electrons in a perpendicular magnetic field.

Medical Physics 2019 January 22
PURPOSE: To evaluate the potential skin dose toxicity contribution of spiralling contaminant electrons (SCE) generated in the air in an MR-linac with a 0.35 T or 1.5 T magnetic field using the EGSnrc Monte Carlo (MC) code. Comparisons to experimental results at 1.5 T are also performed.

METHODS: An Elekta generated phase-space file for the Unity MR-linac is used in conjunction with the EGSnrc enhanced electric and magnetic field transport macros to simulate surface dose profiles and depth dose curves in panels located 5 cm away from the beam edge and positioned either parallel or perpendicular to the magnetic field. Electrons generated in the air will spiral along the magnetic field lines, and though surface doses within the field will be reduced, the electrons can contribute to out-of-field surface doses.

RESULTS: Surface dose profiles showed good agreement with experimental findings and the maximum simulated doses at surfaces perpendicular to the magnetic field were 3.77 ± 0.01% and 3.55 ± 0.01% for 1.5 T and 0.35 T. These results are expressed as a percentage of the maximum dose to water delivered by the photon beam. The surface dose variations in the out-of-field region converge to the 0 T doses within the first 0.5 cm of material. An asymmetry in the dose distribution in surfaces positioned on either side of the photon beam and aligned parallel to the magnetic field is determined to be due to the magnetic field directing electrons deeper into, or localizing them to the surface of, the measurement panel.

CONCLUSIONS: These results confirm the SCE dose contribution in surfaces perpendicular to the magnetic field and show these doses to be of the order of a few percent of the maximum dose to water of the beam. Good agreement in the dose profiles are seen in comparisons between the MC simulations and experimental work. The effect is apparent in 0.35 T and 1.5 T magnetic fields and dissipates within the first few millimetres of material. It should be noted that only SCEs from beam anteriorly incident on the patient will inuence the patient surface dose, and the use of beams incident over different angles will reduce the dose to any particular patient surface. This article is protected by copyright. All rights reserved.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app