Self-measurement with Icare HOME tonometer, patients' feasibility and acceptability.

PURPOSE: To evaluate and compare the accuracy of self-measurement of intraocular pressure using Icare Home rebound tonometer with Goldmann applanation tonometer and assess acceptability of self-tonometry in patients with glaucoma and ocular hypertension.

METHODS: In the study, 117 subjects were trained to use Icare Home for self-measurement. Icare Home tonometer readings were compared with Goldmann applanation tonometer, including one eye per patient. Agreement between the two methods of measurement was evaluated by Bland and Altmann analysis. Questionnaire was used to evaluate patients' perception of self-tonometry.

RESULTS: One hundred and three out of 117 patients (88%) were able to measure their own intraocular pressure and 96 (82%) fulfilled the requirements for certification. The mean (SD) difference Goldmann applanation tonometer minus Icare Home was 1.2 (2.4) mmHg (95% limits of agreement, -3.4 to 5.9 mmHg). The magnitude of bias between the two methods depended on central corneal thickness, with greater bias at central corneal thickness <500 µm. In 65 out of 96 subjects (67.7%), Icare Home results were within 2 mmHg of the Goldmann applanation tonometer. Seventy-three out of 93 (78.5%) felt that self-tonometry was easy to use and 75 patients (80.6%) responded that they would use the device at home.

CONCLUSION: Icare Home tonometry tends to slightly underestimate intraocular pressure compared to Goldmann applanation tonometer. Most patients were able to perform self-tonometry and found it acceptable for home use. Measurements using rebound self-tonometry could improve the quality of intraocular pressure data and optimize treatment regimen.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app