Evaluation of maxillary trabecular microstructure as an indicator of implant stability by using 2 cone beam computed tomography systems and micro-computed tomography.
Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology 2018 December 2
OBJECTIVE: The aim of this study was to assess the trabecular microarchitecture of the maxilla by using cone beam computed tomography (CBCT) and micro-computed tomography (micro-CT) ex vivo.
STUDY DESIGN: Seventeen maxillary cadaver specimens were scanned by using micro-CT and CBCT devices. Samples were scanned with 2 CBCT devices at different voxel sizes (0.08, 0.125, and 0.160 mm for 3-D Accuitomo 170; 0.75 and 0.200 mm for Planmeca Promax 3-D Max). Morphometric parameters, such as bone volume/total volume (BV/TV) ratio, trabecular thickness (Tb.Th), trabecular separation (Tb.Sp), trabecular number (Tb.N), and degree of anisotropy (DA) were assessed by using CTAnalyzer software. Bland-Altman limits of agreement and intraclass correlation coefficient (ICC) were performed to evaluate agreement between CBCT and micro-CT in consideration of measured morphometric parameters. Statistical significance was set at P < .05.
RESULTS: The BV/TV, Tb.Th, Tb.Sp, and DA values were higher for CBCT images compared with micro-CT images, whereas the Tb.N value was lower with CBCT images than with micro-CT images. The BV/TV and DA parameters showed the highest agreement between CBCT and micro-CT devices (ICC = 0.421 for BV/TV and ICC = 0.439 for DA; P < .01).
CONCLUSIONS: The BV/TV and DA parameters measured on CBCT obtained at the smallest voxel size were found to be useful for the assessment of maxillary trabecular microstructure.
STUDY DESIGN: Seventeen maxillary cadaver specimens were scanned by using micro-CT and CBCT devices. Samples were scanned with 2 CBCT devices at different voxel sizes (0.08, 0.125, and 0.160 mm for 3-D Accuitomo 170; 0.75 and 0.200 mm for Planmeca Promax 3-D Max). Morphometric parameters, such as bone volume/total volume (BV/TV) ratio, trabecular thickness (Tb.Th), trabecular separation (Tb.Sp), trabecular number (Tb.N), and degree of anisotropy (DA) were assessed by using CTAnalyzer software. Bland-Altman limits of agreement and intraclass correlation coefficient (ICC) were performed to evaluate agreement between CBCT and micro-CT in consideration of measured morphometric parameters. Statistical significance was set at P < .05.
RESULTS: The BV/TV, Tb.Th, Tb.Sp, and DA values were higher for CBCT images compared with micro-CT images, whereas the Tb.N value was lower with CBCT images than with micro-CT images. The BV/TV and DA parameters showed the highest agreement between CBCT and micro-CT devices (ICC = 0.421 for BV/TV and ICC = 0.439 for DA; P < .01).
CONCLUSIONS: The BV/TV and DA parameters measured on CBCT obtained at the smallest voxel size were found to be useful for the assessment of maxillary trabecular microstructure.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app