Measurement of knee joint range of motion with a digital goniometer: A reliability study.

OBJECTIVES: Measurements of joint range of motion (ROM) are part of a physical therapist's daily work. Activities of daily living and exercises can be complicated to perform when ROM is limited, and depending on the demands in daily living, the knee joint requires different ROM. In sports, a few degrees in ROM may make the difference between getting injured or not. The goals for physical therapists are to help the patients to regain full ROM, mobility, strength, and function after sustaining an injury. To measure joints with the manual universal goniometer is considered time-consuming and difficult with respect to repeated measurements. Recently, a new digital instrument for measuring ROM was developed-EasyAngle. A first objective of the study was to investigate the reliability of EA for measuring knee joint angles, considering intrarater and interrater reliability. A second objective was to investigate if there were any differences in the intrarater reliability between a novice and an experienced assessor.

METHOD: Passive knee angles were measured in fixed positions for 40 knee joints (20 subjects). Two registered physical therapists, one novice and one experienced, conducted the measurements. Both registered physical therapists were blinded to the measurements throughout the study.

RESULTS: The results showed very good intrarater (intraclass correlation coefficient [ICC] 0,997-0,998, standard error of mean 1.15-1.48, smallest detectable difference [SDD] 3.19-4.09, limits of agreement -3.36-3.04, -4.66-4.09) and interrater reliability (ICC 0.994, standard error of mean 2.11, SDD 5.85, limits of agreement -4.75-6.95) for measurements of knee joint ROM. No statistical difference between a novice and an experienced assessor was detected (p = 0.86).

CONCLUSION: The results of the present study showed very good ICC values for both intrarater and interrater reliability measuring knee joint ROM with EasyAngle. Relatively high SDD values were seen for both assessors and may indicate a problem monitoring small differences between measurements. Further studies are recommended to increase the generalizability of the results.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app