A Novel Toroidal-Flow Left Ventricular Assist Device Minimizes Blood Trauma: Implications of Improved Ventricular Assist Device Hemocompatibility.

BACKGROUND: Continuous-flow left ventricular assist devices (LVADs) cause blood trauma that includes von Willebrand factor degradation, platelet activation, and subclinical hemolysis. Blood trauma contributes to bleeding, thrombosis, and stroke, which cause significant morbidity and mortality. The TORVAD (Windmill Cardiovascular Systems, Inc, Austin, TX) is a first-of-its kind, toroidal-flow LVAD designed to minimize blood trauma. We tested the hypothesis that the TORVAD causes less blood trauma than the HeartMate II (Abbott Laboratories, Pleasanton, CA) LVAD.

METHODS: Whole human blood was circulated for 6 hours in ex vivo circulatory loops with a HeartMate II (n = 8; 10,000 rpm, 70 ± 6 mm Hg, 4.0 ± 0.1 L/min) or TORVAD (n = 6; 144 rpm, 72 ± 0.0 mm Hg, 4.3 ± 0.0 L/min). von Willebrand factor degradation was quantified with electrophoresis and immunoblotting. Platelet activation was quantified by cluster of differentiation (CD) 41/61 enzyme-linked immunosorbent assay (ELISA). Hemolysis was quantified by plasma free hemoglobin ELISA.

RESULTS: The TORVAD caused significantly less degradation of high-molecular-weight von Willebrand factor multimers (-10% ± 1% vs -21% ± 1%, p < 0.0001), accumulation of low-molecular-weight von Willebrand factor multimers (22% ± 2% vs 45% ± 2%, p < 0.0001), and accumulation of von Willebrand factor degradation fragments (7% ± 1% vs 25% ± 6%, p < 0.05) than the HeartMate II. The TORVAD did not activate platelets, whereas the HeartMate II caused significant platelet activation (CD 41/61: 645 ± 20 ng/mL vs 1,581 ± 150 ng/mL, p < 0.001; normal human CD 41/61, 593 ng/mL; range, 400 to 800 ng/mL). Similarly, the TORVAD caused minimal hemolysis, whereas the HeartMate II caused significant hemolysis (plasma free hemoglobin: 11 ± 2 vs 109 ± 10 mg/dL, p < 0.0001; normal human plasma free hemoglobin <4 mg/dL).

CONCLUSIONS: The TORVAD design, with markedly lower shear stress and pulsatile flow, caused significantly less blood trauma than the HeartMate II. LVADs with reduced blood trauma are likely to improve clinical outcomes and expand LVAD therapy into patients with less advanced heart failure.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app