BACKGROUND: Fixed-ratio and slope-based algorithms are used to derive oscillometric blood pressure (BP). However, a paucity of published data exists assessing the accuracy of these methods. Our objective was to determine the accuracy of fixed-ratio and slope-based algorithms in healthy adults and in adults with cardiovascular risk factors.
PATIENTS AND METHODS: Overall, 85 healthy adults (age≥18 years) and 85 adults with cardiovascular risk factors were studied. Three oscillometric and four two-observer mercury-based auscultation measurements were performed in each, according to International Standards Organization 2013 methodology. Two fixed-ratio algorithms and one slope-based algorithm were applied to process oscillometric waveform envelopes and derive oscillometric BP. Paired and unpaired t-tests were used to compare mean oscillometric BP within and between each group, respectively.
RESULTS: For healthy adults, mean age was 50.3±17.8 years, mean arm circumference was 30.4±3.8 cm, and 62% were female. In the cardiovascular risk group, mean age was 63.8±12.4 years, mean arm circumference was 31.9±4.2 cm, and 62% were female. For systolic BP, the fixed-ratio algorithms produced the lowest mean error and narrowest SD. For diastolic BP, mean errors were similar for all three algorithms, but the fixed-ratio algorithms had higher precision. The comparison of healthy adults and those with cardiovascular risk factor showed high variability for systolic and diastolic BP (SD: 8.113.9 mmHg).
CONCLUSION: In both healthy adults and in those with cardiovascular risk factors, the fixed-ratio technique performed better than the slope-based algorithm. High between-group variability indicates that subject-specific algorithms may be needed.
PATIENTS AND METHODS: Overall, 85 healthy adults (age≥18 years) and 85 adults with cardiovascular risk factors were studied. Three oscillometric and four two-observer mercury-based auscultation measurements were performed in each, according to International Standards Organization 2013 methodology. Two fixed-ratio algorithms and one slope-based algorithm were applied to process oscillometric waveform envelopes and derive oscillometric BP. Paired and unpaired t-tests were used to compare mean oscillometric BP within and between each group, respectively.
RESULTS: For healthy adults, mean age was 50.3±17.8 years, mean arm circumference was 30.4±3.8 cm, and 62% were female. In the cardiovascular risk group, mean age was 63.8±12.4 years, mean arm circumference was 31.9±4.2 cm, and 62% were female. For systolic BP, the fixed-ratio algorithms produced the lowest mean error and narrowest SD. For diastolic BP, mean errors were similar for all three algorithms, but the fixed-ratio algorithms had higher precision. The comparison of healthy adults and those with cardiovascular risk factor showed high variability for systolic and diastolic BP (SD: 8.113.9 mmHg).
CONCLUSION: In both healthy adults and in those with cardiovascular risk factors, the fixed-ratio technique performed better than the slope-based algorithm. High between-group variability indicates that subject-specific algorithms may be needed.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app