Early failure of aortic bioprostheses: what are the mechanisms?
Current Opinion in Cardiology 2019 March
PURPOSE OF REVIEW: The purpose of this study is to review the contemporary evidence surrounding aortic bioprosthetic valve deterioration, with a focus on early failure of surgically implanted valves.
RECENT FINDINGS: Structural valve deterioration (SVD) remains the most frequent cause of premature bioprosthetic aortic valve failure. However, recent evidence suggests that SVD represents a spectrum, and that clinically silent hemodynamic valve deterioration frequently precedes and predisposes to overt SVD. Hemodynamic valve deterioration is defined as an increase in mean transprosthetic gradient and/or worsening transprosthetic regurgitation on echocardiography. Novel evidence suggests that a dysmetabolic profile may predispose to this phenomenon. Furthermore, subclinical leaflet thrombosis is increasingly recognized as a potential cause of hemodynamic deterioration of bioprosthetic valves. Collectively, these findings highlight the importance of systematic and regular imaging surveillance following bioprosthetic aortic valve replacement.
SUMMARY: Early failure of bioprosthetic aortic valves is a complex and multifactorial phenomenon. Further studies are needed to determine the optimal timing and imaging modality for surveillance following bioprosthetic aortic valve replacement, and to establish strategies to prevent and treat aortic bioprosthetic failure.
RECENT FINDINGS: Structural valve deterioration (SVD) remains the most frequent cause of premature bioprosthetic aortic valve failure. However, recent evidence suggests that SVD represents a spectrum, and that clinically silent hemodynamic valve deterioration frequently precedes and predisposes to overt SVD. Hemodynamic valve deterioration is defined as an increase in mean transprosthetic gradient and/or worsening transprosthetic regurgitation on echocardiography. Novel evidence suggests that a dysmetabolic profile may predispose to this phenomenon. Furthermore, subclinical leaflet thrombosis is increasingly recognized as a potential cause of hemodynamic deterioration of bioprosthetic valves. Collectively, these findings highlight the importance of systematic and regular imaging surveillance following bioprosthetic aortic valve replacement.
SUMMARY: Early failure of bioprosthetic aortic valves is a complex and multifactorial phenomenon. Further studies are needed to determine the optimal timing and imaging modality for surveillance following bioprosthetic aortic valve replacement, and to establish strategies to prevent and treat aortic bioprosthetic failure.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app