JOURNAL ARTICLE
MULTICENTER STUDY
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Surgical Atrioventricular Valve Replacement With Melody Valve in Infants and Children.

Background Pediatric patients with atrioventricular valve disease have limited options for prosthetic valve replacement in sizes <15 mm. Based on successful experience with the stented bovine jugular vein graft (Melody valve) in the right ventricular outflow tract, the prosthesis has been modified for surgical valve replacement in pediatric patients with atrioventricular dysfunction with the intention of subsequent valve expansion in the catheterization laboratory as the child grows. Methods and Results A multicenter, retrospective cohort study was performed among patients who underwent atrioventricular valve replacement with Melody valve at 17 participating sites from North America and Europe, including 68 patients with either mitral (n=59) or tricuspid (n=9) replacement at a median age of 8 months (range, 3 days to 13 years). The median size at implantation was 14 mm (range, 9-24 mm). Immediately postoperatively, the valve was competent with low gradients in all patients. Fifteen patients died; 3 patients underwent transplantation. Nineteen patients required reoperation for adverse outcomes, including valve explantation (n=16), left ventricular outflow tract obstruction (n=1), permanent pacemaker implantation (n=1), and paravalvular leak repair (n=1). Twenty-five patients underwent 41 episodes of catheter-based balloon expansion, exhibiting a significant decrease in median gradient ( P<0.001) with no significant increase in grade of regurgitation. Twelve months after implantation, cumulative incidence analysis indicated that 55% of the patients would be expected to be free from death, heart transplantation, structural valve deterioration, or valve replacement. Conclusions The Melody valve is a feasible option for surgical atrioventricular valve replacement in patients with hypoplastic annuli. The prosthesis shows acceptable short-term function and is amenable to catheter-based enlargement as the child grows. However, patients remain at risk for mortality and structural valve deterioration, despite adequate early valvular function. Device design and implantation techniques must be refined to reduce complications and extend durability. Clinical Trial Registration URL: https://www.clinicaltrials.gov. Unique identifier: NCT02505074.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app