Evaluation of an on-site test device for the heroin metabolite 6-acetylmorphine in urine.

Detection of heroin use is an important task in clinical drug testing and can be best performed by using 6-acetylmorphine as the target analyte. This study was performed to evaluate an on-site test for 6-acetylmorphine screening in urine with an assigned cut-off limit at 10 ng/mL. The reference method was a forensic accredited liquid chromatography-tandem mass spectrometry method. The study confirmed that negative controls and negative authentic specimen resulted in negative readings. Low cross-reactivity was recorded from other potential interfering opioids. Prepared standards and commercial calibrators demonstrated that the cutoff level of the test was lower than the assigned value and rather 2 ng/mL. A study using authentic specimens from patients on substitution treatment with methadone, morphine, and buprenorphine confirmed that the real cut-off level was 2 ng/mL. Using this value as cutoff limit the sensitivity and specificity of the test was 100%.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app