Add like
Add dislike
Add to saved papers

Prospective US investigational device exemption trial of a sutureless aortic bioprosthesis: One-year outcomes.

OBJECTIVES: We performed a prospective, single-arm clinical trial approved under a Food and Drug Administration Investigational Device Exemption to assess safety and efficacy of Perceval, a sutureless bovine pericardial aortic valve representing the initial US experience.

METHODS: From June 2013 to January 2015, 300 patients (mean age 76.7 ± 7.7 years, 54.3% men, 37.3% ≥80 years, median Society of Thoracic Surgeons Predicted Risk of Mortality 2.8%), underwent Perceval valve implantation at 18 centers across the United States. Twenty patients (6.7%) had a bicuspid aortic valve and 5 (1.7%) patients had previous aortic valve replacement. A minimally invasive approach was used in 80 (26.7%) and concomitant procedures were performed in 113 (37.8%) patients.

RESULTS: Two hundred eighty-nine patients (96.3%) were successfully implanted. Operative mortality (≤30 days) was 1.3% (n = 4) (observed to expected ratio of 0.40). One-year results included all-cause mortality in 5.2% (n = 15), stroke in 1% (n = 3), and endocarditis in 1.7% (n = 5). New permanent periprocedural pacemaker rate was 10.7% (n = 30/281); 2.5% (n = 7/281) resulted from third-degree atrioventricular block. One-year valve-related reoperation was 2.1% (n = 6). At 1-year follow-up, 98% of patients were in New York Heart Association class I/II, left ventricular mass index decreased from 103.5 ± 30.1 g/m2 at discharge to 95.8 ± 27.1 g/m2 (P = .001), and 3 (1.3%) moderate paravalvular leaks were identified. Health-related quality of life score increased from 62.7 ± 21.8 before surgery to 85.5 ± 17.8 at 1 year (P < .001).

CONCLUSIONS: These results confirm the safety and effectiveness of the Perceval sutureless aortic valve replacement in study patients with lower mortality than expected from a risk prediction model. Persistent hemodynamic benefit and improvement in quality of life at 1 year support the importance of this device in the management of aortic valve disease.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app