Accuracy of Scheimpflug-derived corneal biomechanical and tomographic indices for detecting subclinical and mild keratectasia in a South Asian population.
Journal of Cataract and Refractive Surgery 2019 March
PURPOSE: To test the predictive accuracy of the Belin-Ambrósio deviation index (BAD-D), the stiffness parameter A1 (SPA1 ), the Corvis biomechanical index (CBI), and the tomographic and biomechanical index (TBI) assessments for distinguishing subclinical and mild keratoconic eyes from normal eyes.
SETTING: Medical Research Foundation, Sankara Nethralaya, Chennai, India.
DESIGN: Retrospective case series.
METHOD: In this cross-sectional clinical study, the following 3 groups of eyes were analyzed: very asymmetric ectasia with normal topography (very asymmetric-normal topography), mild keratoconus, and normal control. All eyes had comprehensive assessment with corneal topography (TMS-IV), Scheimpflug tomography (Pentacam HR), and dynamic Scheimpflug biomechanical analysis (Corvis ST). The outcome measures were the BAD-D, SPA1 , CBI, and TBI. Receiver operating characteristic (ROC) curve analysis was performed to determine each parameter's predictive accuracy in distinguishing between eyes with subclinical or mild keratoconus and control eyes.
RESULTS: The area under the curve (AUC) ROC values for the very asymmetric-normal topography and normal control comparison were 0.81 (BAD-D), 0.76 (SPA1 ), 0.78 (CBI), and 0.90 (TBI). The TBI (using cutoff value 0.16) showed the highest diagnostic accuracy (85%), with 84% sensitivity and 86% specificity. The AUC ROC values for the mild keratoconus and normal control comparison were 0.998 (BAD-D), 0.91 (SPA1 ), 0.97 (CBI), and 0.999 (TBI). The TBI (with a 0.63 cutoff) showed the highest accuracy (99.5%), with 99% sensitivity and 100% specificity. The TBI also showed the weakest correlation with mean keratometry, biomechanically corrected intraocular pressure, and central corneal thickness in normal eyes and keratoconic eyes.
CONCLUSIONS: The TBI best distinguished eyes with mild ectasia from normal eyes and had the weakest correlation with biomechanical confounding factors, reinforcing the hypothesis that the TBI represents corneal biomechanical susceptibility.
SETTING: Medical Research Foundation, Sankara Nethralaya, Chennai, India.
DESIGN: Retrospective case series.
METHOD: In this cross-sectional clinical study, the following 3 groups of eyes were analyzed: very asymmetric ectasia with normal topography (very asymmetric-normal topography), mild keratoconus, and normal control. All eyes had comprehensive assessment with corneal topography (TMS-IV), Scheimpflug tomography (Pentacam HR), and dynamic Scheimpflug biomechanical analysis (Corvis ST). The outcome measures were the BAD-D, SPA1 , CBI, and TBI. Receiver operating characteristic (ROC) curve analysis was performed to determine each parameter's predictive accuracy in distinguishing between eyes with subclinical or mild keratoconus and control eyes.
RESULTS: The area under the curve (AUC) ROC values for the very asymmetric-normal topography and normal control comparison were 0.81 (BAD-D), 0.76 (SPA1 ), 0.78 (CBI), and 0.90 (TBI). The TBI (using cutoff value 0.16) showed the highest diagnostic accuracy (85%), with 84% sensitivity and 86% specificity. The AUC ROC values for the mild keratoconus and normal control comparison were 0.998 (BAD-D), 0.91 (SPA1 ), 0.97 (CBI), and 0.999 (TBI). The TBI (with a 0.63 cutoff) showed the highest accuracy (99.5%), with 99% sensitivity and 100% specificity. The TBI also showed the weakest correlation with mean keratometry, biomechanically corrected intraocular pressure, and central corneal thickness in normal eyes and keratoconic eyes.
CONCLUSIONS: The TBI best distinguished eyes with mild ectasia from normal eyes and had the weakest correlation with biomechanical confounding factors, reinforcing the hypothesis that the TBI represents corneal biomechanical susceptibility.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app