Reliability of the EOS Imaging System for Assessment of the Spinal and Pelvic Alignment in the Sagittal Plane.
Clinics in Orthopedic Surgery 2018 December
Background: The sagittal alignment of the spine and pelvis is not only closely related to the overall posture of the body but also to the evaluation and treatment of spine disease. In the last few years, the EOS imaging system, a new low-dose radiation X-ray device, became available for sagittal alignment assessment. However, there has been little research on the reliability of EOS. The purpose of this study was to evaluate the intrarater and interrater reliability of EOS for the sagittal alignment assessment of the spine and pelvis.
Methods: Records of 46 patients were selected from the EOS recording system between November 2016 and April 2017. The exclusion criteria were congenital spinal anomaly and deformity, and previous history of spine and pelvis operation. Sagittal parameters of the spine and pelvis were measured by three examiners three times each using both manual and EOS methods. Means comparison t -test, Pearson bivariate correlation analysis, and reliability analysis by intraclass correlation coefficients (ICCs) for intrarater and interrater reliability were performed using R package "irr."
Results: We found excellent intrarater and interrater reliability of EOS measurements. For intrarater reliability, the ICC ranged from 0.898 to 0.982. For interrater reliability, the ICC ranged from 0.794 to 0.837. We used a paired t -test to compare the values measured by manual and EOS methods: there was no statistically significant difference between the two methods. Correlation analysis also showed a statistically significant positive correlation.
Conclusions: EOS showed excellent reliability for assessment of the sagittal alignment of the spine and pelvis.
Methods: Records of 46 patients were selected from the EOS recording system between November 2016 and April 2017. The exclusion criteria were congenital spinal anomaly and deformity, and previous history of spine and pelvis operation. Sagittal parameters of the spine and pelvis were measured by three examiners three times each using both manual and EOS methods. Means comparison t -test, Pearson bivariate correlation analysis, and reliability analysis by intraclass correlation coefficients (ICCs) for intrarater and interrater reliability were performed using R package "irr."
Results: We found excellent intrarater and interrater reliability of EOS measurements. For intrarater reliability, the ICC ranged from 0.898 to 0.982. For interrater reliability, the ICC ranged from 0.794 to 0.837. We used a paired t -test to compare the values measured by manual and EOS methods: there was no statistically significant difference between the two methods. Correlation analysis also showed a statistically significant positive correlation.
Conclusions: EOS showed excellent reliability for assessment of the sagittal alignment of the spine and pelvis.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
The Effect of Albumin Administration in Critically Ill Patients: A Retrospective Single-Center Analysis.Critical Care Medicine 2024 Februrary 8
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app