A novel signal-averaged electrocardiogram and an ambulatory-based signal-averaged electrocardiogram show strong correlations with conventional signal-averaged electrocardiogram in healthy subjects: A validation study.

BACKGROUND: A novel signal-averaged electrocardiogram (SAECG) device and a novel ambulatory SAECG device are clinically available, but reference values have not been established. This study aimed to validate the novel SAECG and the novel ambulatory-based SAECG devices by comparison with the conventional SAECG device.

METHODS AND RESULTS: High-resolution SAECGs were recorded consecutively in 83 healthy volunteers using the 3 devices. A novel ambulatory SAECG device was used as real-time recording within 15 min for validation study (15 min ambulatory-based SAECG). We examined the concordance of positive results (at least 2/3 abnormal SAECG parameters) and negative results (0 or 1/3 abnormal parameters), as well as the correlations between SAECG parameters (filtered QRS duration [fQRS]); duration of low-amplitude signals < 40 μV in the terminal filtered QRS complex [LAS40 ]; root mean square voltage of the terminal 40 ms of the filtered QRS complex [RMS40 ]). Qualitative analysis showed excellent concordance among the novel SAECG, the 15 min ambulatory-based SAECG, and the conventional SAECG methods (novel SAECG vs. conventional SAECG = 94%; 15 min ambulatory-based SAECG vs. conventional SAECG = 91.6%; p = 0.755), while quantitative analysis indicated strong correlations between the novel SAECG and the conventional SAECG values for fQRS, LAS40 , and LnRMS40 (r = 0.838-0.805, p < 0.0001, respectively). Strong correlations were also seen between 15 min ambulatory-based SAECG and conventional SAECG values for fQRS, LAS40 , and RMS40 (r = 0.943-0.888, p < 0.0001, respectively). However, Bland-Altman quantitative analysis showed better agreement in fQRS and LnRMS40 measured by the 15 min ambulatory-based SAECG and the conventional SAECG than those by the novel SAECG and the conventional SAECG (fQRS, Lin's rho_c = 0.923 vs. 0757; RMS40 , Lin's rho_c = 0.932 vs. 0.818, respectively).

CONCLUSION: In healthy subjects, the parameters of either the novel SAECG or the 15 min ambulatory-based SAECG and those of the conventional SAECG were strongly correlated. Relatively good agreements were observed among 3 SAECGs, especially better between the 15 min ambulatory-based SAECG and the conventional SAECG probably due to similar measurement system of 2 methods.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app