Reliability and Agreement of Various InBody Body Composition Analyzers as Compared to Dual-Energy X-Ray Absorptiometry in Healthy Men and Women.
BACKGROUND: Bioelectrical impedance analysis has evolved over the years to include the use of multiple frequencies and impedance measurements to improve the accuracy and reliability of body composition estimates. The purpose of this investigation was to evaluate the reliability of the InBody230 , InBody720 , and InBody770 to measure body fat percent (BF%), fat mass (FM), and fat-free mass (FFM) in the general population and to compare results to dual-energy X-ray absorptiometry (DXA).
METHODS: A total of 31 males and 36 females participated in 2 d of testing separated by 24-72 h. Each visit consisted of a DXA scan, and analysis with the InBody230 , InBody720 , and InBody770 .
RESULTS: All 3 bioelectrical impedance devices (InBody230 , InBody720 , and InBody770 ) were reliable in men and women as indicated by high intraclass correlation coefficients for BF% (≥0.98), FM (≥0.98), and FFM (≥0.99) and low standard error of measurement for BF% (0.77%-0.99%), FM (0.54-0.87 kg), and FFM (0.58-0.84 kg) and minimum difference for BF% (2.12%-2.73%), FM (1.49-2.39 kg), and FFM (1.60-2.32 kg), respectively. When examining the agreement between the 3 InBody analyzers with DXA, systematic bias (underestimation of BF% and FM and overestimation of FFM) was present for all comparisons (p < 0.05) while proportional bias was present for FM in women and FFM in men. However, there was small individual error for all comparisons as indicated by the standard error of estimate and 95% limits of agreement.
CONCLUSION: The InBody analyzers produce small individual error, which suggest these methods can be used as a surrogate when DXA is not available; however, practitioners should be aware of the systematic bias for all comparisons and proportional bias for FM in women and FFM in men. Furthermore, findings revealed that the research grade models, InBody720 and InBody770 , added minimal benefit over the portable InBody230 when assessing BF%, FM, and FFM.
METHODS: A total of 31 males and 36 females participated in 2 d of testing separated by 24-72 h. Each visit consisted of a DXA scan, and analysis with the InBody230 , InBody720 , and InBody770 .
RESULTS: All 3 bioelectrical impedance devices (InBody230 , InBody720 , and InBody770 ) were reliable in men and women as indicated by high intraclass correlation coefficients for BF% (≥0.98), FM (≥0.98), and FFM (≥0.99) and low standard error of measurement for BF% (0.77%-0.99%), FM (0.54-0.87 kg), and FFM (0.58-0.84 kg) and minimum difference for BF% (2.12%-2.73%), FM (1.49-2.39 kg), and FFM (1.60-2.32 kg), respectively. When examining the agreement between the 3 InBody analyzers with DXA, systematic bias (underestimation of BF% and FM and overestimation of FFM) was present for all comparisons (p < 0.05) while proportional bias was present for FM in women and FFM in men. However, there was small individual error for all comparisons as indicated by the standard error of estimate and 95% limits of agreement.
CONCLUSION: The InBody analyzers produce small individual error, which suggest these methods can be used as a surrogate when DXA is not available; however, practitioners should be aware of the systematic bias for all comparisons and proportional bias for FM in women and FFM in men. Furthermore, findings revealed that the research grade models, InBody720 and InBody770 , added minimal benefit over the portable InBody230 when assessing BF%, FM, and FFM.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app