Modified ERG-Jet™ contact lens electrodes for use in infants and toddlers-update.
Documenta Ophthalmologica. Advances in Ophthalmology 2018 December
PURPOSE: To describe recent changes in the dimensions of the ERG-jet contact lens electrode, and to suggest corresponding changes in the dimensions of cylindrical stents to facilitate the use of ERG-jet contact lens electrodes in infants and small children.
METHODS: Upon the observation that existing cylindrical stents no longer fit newly purchased ERG-jet contact lenses, the spacings between the "bumps" on the front surface of old and new electrodes were measured with calipers, and prototype stents were newly machined from acrylic plastic.
RESULTS: Cylindrical stents with an outside diameter of 0.277 inches (about 7.04 mm) were found to fit snugly between the front surface "bumps" of new ERG-jet contact lens electrodes.
CONCLUSIONS: Recent modifications in the injection molds for ERG-jet contact lens electrodes have resulted in an increase in the space between the "bumps" on the front surface of the electrodes, necessitating a corresponding increase in the outside diameter of cylindrical stents intended to fit snugly between the "bumps" for ease of insertion and prevention of lid closure.
METHODS: Upon the observation that existing cylindrical stents no longer fit newly purchased ERG-jet contact lenses, the spacings between the "bumps" on the front surface of old and new electrodes were measured with calipers, and prototype stents were newly machined from acrylic plastic.
RESULTS: Cylindrical stents with an outside diameter of 0.277 inches (about 7.04 mm) were found to fit snugly between the front surface "bumps" of new ERG-jet contact lens electrodes.
CONCLUSIONS: Recent modifications in the injection molds for ERG-jet contact lens electrodes have resulted in an increase in the space between the "bumps" on the front surface of the electrodes, necessitating a corresponding increase in the outside diameter of cylindrical stents intended to fit snugly between the "bumps" for ease of insertion and prevention of lid closure.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app