We have located links that may give you full text access.
Development of Repeatable Microcatheter Access Port for Intra-arterial Therapy of Liver Cancer.
Cardiovascular and Interventional Radiology 2018 November 3
PURPOSE: To develop an implantable port in which a microcatheter can be inserted for a combination therapy of repeated transarterial chemoembolization (TACE) and hepatic arterial infusion chemotherapy (HAIC) for advanced liver cancer.
MATERIALS AND METHODS: The design of a currently used implantable port was modified. A funnel part was constructed in the port. The septum was punctured by a 20-gauge indwelling needle, and 2.0-Fr non-tapered microcatheter was inserted into the port. In the in vitro studies, the advance of a microcatheter out of the funnel part was evaluated via seven different septum puncture sites. A 5-Fr indwelling catheter connected to the port was placed in a vascular model, and a microcatheter catheterization was evaluated. In an in vivo study, the port-catheter system was implanted in the hepatic artery in a pig. A microcatheter was percutaneously inserted through the port into the hepatic arterial branches, and embolization was performed.
RESULTS: In the in vitro studies, the microcatheter was smoothly advanced out of the port and catheterizations into the hepatic arteries were successful via all septum puncture sites. In the in vivo study, repeated selective embolization through the port was successfully conducted on 7, 14 and 21 days after the implantation.
CONCLUSION: The developed implantable port can be used for repeated catheter insertion into the hepatic artery. The combination of repeated TACE and HAIC could be possible using this device.
MATERIALS AND METHODS: The design of a currently used implantable port was modified. A funnel part was constructed in the port. The septum was punctured by a 20-gauge indwelling needle, and 2.0-Fr non-tapered microcatheter was inserted into the port. In the in vitro studies, the advance of a microcatheter out of the funnel part was evaluated via seven different septum puncture sites. A 5-Fr indwelling catheter connected to the port was placed in a vascular model, and a microcatheter catheterization was evaluated. In an in vivo study, the port-catheter system was implanted in the hepatic artery in a pig. A microcatheter was percutaneously inserted through the port into the hepatic arterial branches, and embolization was performed.
RESULTS: In the in vitro studies, the microcatheter was smoothly advanced out of the port and catheterizations into the hepatic arteries were successful via all septum puncture sites. In the in vivo study, repeated selective embolization through the port was successfully conducted on 7, 14 and 21 days after the implantation.
CONCLUSION: The developed implantable port can be used for repeated catheter insertion into the hepatic artery. The combination of repeated TACE and HAIC could be possible using this device.
Full text links
Related Resources
Trending Papers
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app