We have located links that may give you full text access.
The limitations and potential adverse effects of the premature ventricular contraction response.
Journal of Arrhythmia 2018 October
A 69-year-old man, who had undergone surgery for mitral and tricuspid regurgitation with the Maze procedure for paroxysmal atrial fibrillation, was admitted with an episode of syncope due to sick sinus syndrome. Three days after implantation of a dual-chamber pacemaker (Accent MRI™, St. Jude Medical Inc.,), ventricular pacing on T-wave was recorded multiple times. St. Jude Medical Inc. pacemakers have a unique additional algorithm, called premature ventricular contraction response, related to preventing pacemaker-mediated tachycardia. This algorithm was determined to be a cause of ventricular pacing on T-wave. We report the limitations and potential adverse effects of such automated algorithms.
Full text links
Related Resources
Trending Papers
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app