Add like
Add dislike
Add to saved papers

Development and Validation of a High-Performance Liquid Chromatographic Method for the Determination of Cinitapride in Human Plasma.

A precise and reliable reversed-phase high-performance liquid chromatographic method with ultraviolet detection was developed and validated to determine cinitapride in human plasma. After liquid-liquid extraction, chromatographic separation was achieved on a Nucleosil C18 (25 cm × 4.6 mm, 5  µ m) column with an isocratic elution consisting of 10 mM ammonium acetate (pH 5.2), methanol, and acetonitrile, 40 : 50 : 10, v/v/v. The developed method was validated as per US FDA guidelines for its linearity, selectivity, sensitivity, precision, accuracy, and stability. Satisfactory findings were obtained from the validation studies. The linearity range of the method was 1 to 35 ng/mL while the extraction recovery of cinitapride in human plasma was more than 86%. The percent coefficient of variation of both intraday and interday precision was ≤7.1%.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app