Evaluating a Portable, Noncontact Fundus Camera for Retinopathy of Prematurity Screening by Nonophthalmologist Health Care Workers.
Ophthalmology Retina 2018 August
PURPOSE: To evaluate (1) the feasibility of non-ophthalmologist healthcare workers (HCWs) to obtain images of sufficient quality for retinopathy of prematurity (ROP) screening using a FDA-approved portable, non-contact, narrow-field fundus camera (i.e., Pictor™), and (2) the accuracy of grading these images to identify infants who developed treatment-warranted (type 1) ROP.
DESIGN: Prospective cohort study.
SUBJECTS: Infants undergoing routine ROP screening examinations (i.e. birth weight ≤1500 grams and/or gestational age ≤30 weeks or selected infants with a birth weight of 1500-2000g or gestational age >30 weeks and an unstable clinical course).
METHODS: We prospectively recruited infants undergoing ROP screening examinations at a community hospital. On the same day an ophthalmologist examined them, a trained HCW imaged their retinas using the non-contact camera. Two masked ROP experts graded these images remotely. We calculated both the percentage of gradable images (i.e. having at least 3 quadrants with sufficient image quality), as well as the accuracy of identifying infants who developed type 1 ROP.
MAIN OUTCOME MEASURES: Percentage of gradable images and the sensitivity and specificity of each grader for identifying infants with type 1 ROP by grading for the presence of pre-plus or plus disease.
RESULTS: Ninety-nine infants were included. Overall, 92.4% and 94.2% of all infant imaging sessions were considered gradable by graders 1 and 2, respectively. Amongst gradable images, the sensitivity of both graders for identifying type 1 ROP by grading for the presence of pre-plus or plus disease was 100% (95% confidence interval (CI): 95-100%) and the specificity 91% (95% CI: 83-95%) for grader 1 and 93% (95% CI: 86-96%) for grader 2.
CONCLUSIONS: It was highly feasible for trained HCWs to obtain digital retinal images of sufficient quality for ROP screening using a non-contact fundus camera. By grading for the presence of pre-plus or plus disease, graders identified infants who developed type 1 ROP with high sensitivity and specificity. The use of portable, non-contact retinal cameras by trained HCWs could increase our workforce in ROP screening and identify infants needing an indirect ophthalmoscopy examination by an ophthalmologist.
DESIGN: Prospective cohort study.
SUBJECTS: Infants undergoing routine ROP screening examinations (i.e. birth weight ≤1500 grams and/or gestational age ≤30 weeks or selected infants with a birth weight of 1500-2000g or gestational age >30 weeks and an unstable clinical course).
METHODS: We prospectively recruited infants undergoing ROP screening examinations at a community hospital. On the same day an ophthalmologist examined them, a trained HCW imaged their retinas using the non-contact camera. Two masked ROP experts graded these images remotely. We calculated both the percentage of gradable images (i.e. having at least 3 quadrants with sufficient image quality), as well as the accuracy of identifying infants who developed type 1 ROP.
MAIN OUTCOME MEASURES: Percentage of gradable images and the sensitivity and specificity of each grader for identifying infants with type 1 ROP by grading for the presence of pre-plus or plus disease.
RESULTS: Ninety-nine infants were included. Overall, 92.4% and 94.2% of all infant imaging sessions were considered gradable by graders 1 and 2, respectively. Amongst gradable images, the sensitivity of both graders for identifying type 1 ROP by grading for the presence of pre-plus or plus disease was 100% (95% confidence interval (CI): 95-100%) and the specificity 91% (95% CI: 83-95%) for grader 1 and 93% (95% CI: 86-96%) for grader 2.
CONCLUSIONS: It was highly feasible for trained HCWs to obtain digital retinal images of sufficient quality for ROP screening using a non-contact fundus camera. By grading for the presence of pre-plus or plus disease, graders identified infants who developed type 1 ROP with high sensitivity and specificity. The use of portable, non-contact retinal cameras by trained HCWs could increase our workforce in ROP screening and identify infants needing an indirect ophthalmoscopy examination by an ophthalmologist.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app