Implementing the centiloid transformation for 11 C-PiB and β-amyloid 18 F-PET tracers using CapAIBL.

NeuroImage 2018 August 19
The centiloid scale was recently proposed to provide a standard framework for the quantification of β-amyloid PET images, so that amyloid burden can be expressed on a standard scale. While the framework prescribes SPM8 as the standard analysis method for PET quantification, non-standard methods can be calibrated to produce centiloid values. We have previously developed a PET-only quantification: CapAIBL. In this study, we show how CapAIBL can be calibrated to the centiloid scale.

METHODS: Calibration images for 11 C-PiB, 18 F-NAV4694, 18 F-Florbetaben, 18 F-Flutemetamol and 18 F- Florbetapir were analysed using the standard method and CapAIBL. Using these images, both methods were calibrated to the centiloid scale. Centiloid values computed using CapAIBL were compared to those computed using standard method. For each tracer, a separate validation was performed using an independent dataset from the AIBL study.

RESULTS: Using the calibration images, there was a very strong agreement, and very little bias between the centiloid values computed using CapAIBL and those computed using the standard method with R2  > 0.97 across all tracers. Using images from AIBL, the agreement was also high with R2  > 0.96 across all tracers. In this dataset, there was a small underestimation of the centiloid values computed using CapAIBL of less than 0.8% in PiB, and a small over-estimation of 1.3% in Florbetapir, and 0.8% in Flutemetamol. There was a larger overestimation of 8% in NAV images, and 14% underestimation in Florbetaben images. However, some of these differences could be explained by the use of different scanners between the calibration scans and the ones used in AIBL.

CONCLUSION: The PET-only quantification method, CapAIBL, can produce reliable centiloid values. The bias observed in the AIBL dataset for 18F-NAV4694 and 18 F-Florbetaben may indicate that using different scanners or reconstruction methods might require scanner-specific adjustments.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app