Comparative Study
Evaluation Study
Journal Article
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Evaluation of a Validated Luminex-Based Multiplex Immunoassay for Measuring Immunoglobulin G Antibodies in Serum to Pneumococcal Capsular Polysaccharides.

MSphere 2018 August 9
This article describes the results of a study designed to bridge the World Health Organization (WHO) pneumococcal enzyme-linked immunosorbent assay (ELISA) platform to the validated Luminex-based 13-plex direct immunoassay (dLIA) platform developed by Pfizer, Inc. Both assay platforms quantify serotype-specific serum IgG antibodies (in micrograms per milliliter) against an international reference standard serum. The primary goal of this study was to determine if the dLIA is a suitable replacement for the ELISA to support clinical vaccine studies that include the evaluation of immune responses to serotypes 1, 3, 4, 5, 6A, 6B, 7F, 9V, 14, 18C, 19A, 19F, and 23F. Serum samples were selected from four pivotal 13-valent pneumococcal conjugate vaccine (13vPnC; Prevnar 13) clinical trials on the basis of their serotype-specific IgG concentrations by ELISA. In these studies, subjects were immunized either with 13vPnC or with 7-valent pneumococcal conjugate vaccine (7vPnC; Prevnar). There were 1,528 of 1,574 selected samples with sufficient remaining volume for reanalysis in the dLIA. A comparison of assay results from the dLIA and ELISA platforms showed clear and robust linear quantitative relationships across all 13 serotypes. In addition, lower IgG antibody concentrations in preimmunization samples were measured in the dLIA, thus allowing better differentiation between preimmunization and low-titer postimmunization samples. Overall, the results showed that the established population-level protective threshold IgG concentration, 0.35 µg/ml of serotype-specific serum IgG antibodies, is appropriate for use for data generated using the dLIA platform developed by Pfizer, Inc., for 10 serotypes: serotypes 1, 3, 4, 6A, 7F, 9V, 14, 18C, 19F, and 23F. On the basis of the extensive bridging analyses, however, the use of dLIA cutoff values of 0.23, 0.10, and 0.12 µg/ml is recommended for serotypes 5, 6B, and 19A, respectively. This adjustment will ensure that the consistency of the established population-level protective threshold IgG concentration is maintained when switching from the ELISA to the dLIA platform. The results of this bridging study demonstrate that the 13-plex dLIA platform is a suitable replacement for the WHO reference ELISA platform. IMPORTANCE The pneumococcal enzyme-linked immunosorbent assay (ELISA) measures IgG antibodies in human serum, and it is an important assay that supports licensure of pneumococcal vaccines. The immune correlate of protection, 0.35 µg/ml of IgG antibodies, was determined by the ELISA method. Pfizer has developed a new Luminex-based assay platform to replace the ELISA. These papers describe the important work of (i) validating the Luminex-based assay and (ii) bridging the immune correlate of protection (0.35 µg/ml IgG) to equivalent values reported by the Luminex platform.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app