Add like
Add dislike
Add to saved papers

Small bowel obstruction SBO after TAPP repair caused by a self-anchoring barbed suture device for peritoneal closure: case report.

A 50-year-old man underwent laparoscopic hernia repair for a groin hernia, presenting acute abdominal pain and bowel obstruction syndrome 1 day post surgery. Diagnostic laparoscopy was performed at postoperative the day after the hernioplasty and a volvulus was found. The residual end of the barbed V-LOC adopted in the peritoneal closure was incidentally hooked to the mesentery and caused a small bowel obstruction as a volvulus. The redundant V-LOC strand was released and cut superficial to the peritoneum. A detorsion of volvulus was preformed. Neither bowel ischemia nor significant bowel injury was noted. The following day he was discharged without complication. The residual 'free' barbel suture in the peritoneal cavity invited adhesion formations and subsequently the Bowel obstruction.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app