High-speed recording of thermal load during laser trans-epithelial corneal refractive surgery using a 750Hz ablation system.

PURPOSE: To evaluate the temperature rise of human cornea during trans-epithelial photorefractive keratectomy (trans-PRK) with a 750Hz excimer laser employing Intelligent Thermal Effect Control (ITEC) software.

METHODS: In this observational case series, trans-PRK ablation was performed on 5 eyes of 3 patients using an aspheric profile of a 750Hz excimer laser system. A high-resolution infrared camera with a frame-rate of 350 images per second was used to determine the corneal surface temperature. Images were taken sequentially, starting a few seconds prior to and ending a few seconds after the ablation. The maximum temperature of any pixel of a given image were recorded and graphed against time.

RESULTS: The baseline ocular surface temperature, immediately prior to the beginning of excimer laser, ranged from 32 to 34.9°C. The maximum ocular surface temperature until the epithelium was ablated ranged from 35.2 to 39.7°C. The maximum ocular surface temperature during stromal ablation with high and low fluence laser ranged from 32.9 to 36.5 and from 34.4 to 37.7°C respectively.

CONCLUSION: The ITEC software is effective in controlling the maximum temperature rise during laser ablation in the extremely challenging situation of trans-PRK involving high ablation volumes of almost 6000nl, potentially improving the outcomes. The ITEC system limited the maximum temperature to 39.7°C in the epithelium, and 37.7°C in the stroma. The epithelial temperature was always higher than stromal temperature (regardless of high or low fluence irradiation). Safety limit of 40°C found in the literature was never reached.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app