Adjustable-Loop Femoral Cortical Suspensory Fixation for Patellar Tendon Anterior Cruciate Ligament Reconstruction: A Time Zero Biomechanical Comparison With Interference Screw Fixation.

BACKGROUND: Adjustable-loop cortical buttons for femoral fixation of bone-tendon-bone grafts have potential advantages over interference screw fixation; however, these devices have not been benchmarked biomechanically against interference screws. Purpose/Hypothesis: The purpose was to compare the time zero biomechanical properties of commercially available, adjustable-loop cortical button and metallic interference screws for femoral fixation of bone-tendon-bone grafts. It was hypothesized that no significant differences would be found in biomechanical properties between fixation techniques.

STUDY DESIGN: Controlled laboratory study.

METHODS: Adjustable-loop cortical buttons (n = 8) and metallic interference screws (n = 8) were used to fix matched pairs of human bone-tendon-bone allografts in porcine distal femurs. These constructs were preconditioned (10 N to 50 N at 1 Hz, 10 cycles), subjected to cyclic loading (50 N to 250 N at 1 Hz, 500 cycles), and then pulled to failure at 20 mm/min.

RESULTS: The loads to failure (mean ± SD, 700 ± 256 N vs 688 ± 215 N, P = .92) and linear stiffnesses (219 ± 48 N/mm vs 218 ± 49 N/mm, P = .97) for the adjustable-loop cortical button and metallic interference screws, respectively, were not significantly different. Cyclic displacement was higher in the adjustable-loop cortical button group (2.1 ± 0.6 mm vs 1.3 ± 0.4 mm, P = .01). The mechanism of failure was different between groups, with bone block slippage occurring most commonly in the interference screw group (n = 5) and fracture of the bone block through the suture hole occurring most commonly in the adjustable-loop cortical button group (n = 6).

CONCLUSION: Adjustable-loop cortical buttons and interference screws have similar time zero failure loads, although cyclic displacement was higher with the adjustable-loop cortical buttons. The mean difference in displacement was less than 1 mm compared with the interference screw.

CLINICAL RELEVANCE: Adjustable-loop cortical buttons may be an acceptable alternative to an interference screw for femoral fixation of bone-tendon-bone grafts in anterior cruciate ligament reconstruction. The clinical relevance of the observed differences in cyclic displacement is unknown and should be evaluated in future studies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

Managing Alcohol Withdrawal Syndrome.Annals of Emergency Medicine 2024 March 26

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app