We have located links that may give you full text access.
12-Month Results From the First-in-Human Randomized Study of the Ranger Paclitaxel-Coated Balloon for Femoropopliteal Treatment.
JACC. Cardiovascular Interventions 2018 May 29
OBJECTIVES: The authors sought to evaluate the performance of the Ranger paclitaxel-coated balloon versus uncoated balloon angioplasty for femoropopliteal lesions at 12 months.
BACKGROUND: Drug-coated balloons (DCBs) are a promising endovascular treatment option for peripheral artery disease of the femoropopliteal segment, and each unique device requires dedicated clinical study.
METHODS: The prospective, randomized RANGER SFA (Comparison of the Ranger™ Paclitaxel-Coated PTA Balloon Catheter and Uncoated PTA Balloons in Femoropopliteal Arteries) study (NCT02013193) enrolled 105 patients with symptomatic lower limb ischemia (Rutherford category 2 to 4) and stenotic lesions in the nonstented femoropopliteal segment at 10 European centers. Seventy-one patients (mean age 68 ± 8 years, n = 53 men) were enrolled in the Ranger DCB arm, and 34 patients (mean age 67 ± 9 years, n = 23 men) were assigned to the control group. Twelve-month analysis included patency, safety, and clinical outcomes and quality-of-life assessments.
RESULTS: The DCB group had a greater primary patency rate at 12 months (Kaplan-Meier estimate 86.4% vs. 56.5%), with a significantly longer time to patency failure (log-rank p < 0.001). The estimated freedom from target lesion revascularization rate was 91.2% in the DCB group and 69.9% in the control group at 12 months, with a significantly longer time to reintervention (p = 0.010). No target limb amputations or device-related deaths occurred in either group.
CONCLUSIONS: Twelve-month results show that patency was maintained longer after Ranger DCB treatment than after conventional balloon angioplasty, and this result was associated with a low revascularization rate and good clinical outcomes.
BACKGROUND: Drug-coated balloons (DCBs) are a promising endovascular treatment option for peripheral artery disease of the femoropopliteal segment, and each unique device requires dedicated clinical study.
METHODS: The prospective, randomized RANGER SFA (Comparison of the Ranger™ Paclitaxel-Coated PTA Balloon Catheter and Uncoated PTA Balloons in Femoropopliteal Arteries) study (NCT02013193) enrolled 105 patients with symptomatic lower limb ischemia (Rutherford category 2 to 4) and stenotic lesions in the nonstented femoropopliteal segment at 10 European centers. Seventy-one patients (mean age 68 ± 8 years, n = 53 men) were enrolled in the Ranger DCB arm, and 34 patients (mean age 67 ± 9 years, n = 23 men) were assigned to the control group. Twelve-month analysis included patency, safety, and clinical outcomes and quality-of-life assessments.
RESULTS: The DCB group had a greater primary patency rate at 12 months (Kaplan-Meier estimate 86.4% vs. 56.5%), with a significantly longer time to patency failure (log-rank p < 0.001). The estimated freedom from target lesion revascularization rate was 91.2% in the DCB group and 69.9% in the control group at 12 months, with a significantly longer time to reintervention (p = 0.010). No target limb amputations or device-related deaths occurred in either group.
CONCLUSIONS: Twelve-month results show that patency was maintained longer after Ranger DCB treatment than after conventional balloon angioplasty, and this result was associated with a low revascularization rate and good clinical outcomes.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app