A Kinematic Sensor and Algorithm to Detect Motor Fluctuations in Parkinson Disease: Validation Study Under Real Conditions of Use.
JMIR Rehabilitation and Assistive Technologies 2018 April 26
BACKGROUND: A new algorithm has been developed, which combines information on gait bradykinesia and dyskinesia provided by a single kinematic sensor located on the waist of Parkinson disease (PD) patients to detect motor fluctuations (On- and Off-periods).
OBJECTIVE: The goal of this study was to analyze the accuracy of this algorithm under real conditions of use.
METHODS: This validation study of a motor-fluctuation detection algorithm was conducted on a sample of 23 patients with advanced PD. Patients were asked to wear the kinematic sensor for 1 to 3 days at home, while simultaneously keeping a diary of their On- and Off-periods. During this testing, researchers were not present, and patients continued to carry on their usual daily activities in their natural environment. The algorithm's outputs were compared with the patients' records, which were used as the gold standard.
RESULTS: The algorithm produced 37% more results than the patients' records (671 vs 489). The positive predictive value of the algorithm to detect Off-periods, as compared with the patients' records, was 92% (95% CI 87.33%-97.3%) and the negative predictive value was 94% (95% CI 90.71%-97.1%); the overall classification accuracy was 92.20%.
CONCLUSIONS: The kinematic sensor and the algorithm for detection of motor-fluctuations validated in this study are an accurate and useful tool for monitoring PD patients with difficult-to-control motor fluctuations in the outpatient setting.
OBJECTIVE: The goal of this study was to analyze the accuracy of this algorithm under real conditions of use.
METHODS: This validation study of a motor-fluctuation detection algorithm was conducted on a sample of 23 patients with advanced PD. Patients were asked to wear the kinematic sensor for 1 to 3 days at home, while simultaneously keeping a diary of their On- and Off-periods. During this testing, researchers were not present, and patients continued to carry on their usual daily activities in their natural environment. The algorithm's outputs were compared with the patients' records, which were used as the gold standard.
RESULTS: The algorithm produced 37% more results than the patients' records (671 vs 489). The positive predictive value of the algorithm to detect Off-periods, as compared with the patients' records, was 92% (95% CI 87.33%-97.3%) and the negative predictive value was 94% (95% CI 90.71%-97.1%); the overall classification accuracy was 92.20%.
CONCLUSIONS: The kinematic sensor and the algorithm for detection of motor-fluctuations validated in this study are an accurate and useful tool for monitoring PD patients with difficult-to-control motor fluctuations in the outpatient setting.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app