Valve in valve implantation of the CoreValve Evolut R in degenerated surgical aortic valves.
Cardiology Journal 2018
BACKGROUND: The new CoreValve Evolut R has an improved design to minimize paravalvular leak-age and allows repositioning of the valve. For patients with degenerated bioprosthetic aortic valves, transcatheter aortic valve implantation (TAVI) represents a less invasive option. Herein reported are valve-in-valve (ViV) implantations of this new valve.
METHODS: A total of 26 patients (mean age 79.4 ± 6.1 years, 17 males and 9 females) were treated for severe prosthesis stenosis (n = 9), severe regurgitation (n = 8) or severe combination of stenosis and regurgitation (n = 9). All patients underwent transthoracic echocardiography before and after ViV implantation.
RESULTS: Valve-in-valve implantation of a CoreValve Evolut R was performed successfully in all pa-tients. The mean transaortic gradient for stenotic valves determined by transthoracic echocardiography was reduced significantly from 37.5 ± 15.3 mmHg in patients with prosthesis stenosis to 16.3 ± 8.2 mmHg (p < 0.001). In all cases with severe prosthesis regurgitation, regurgitation was reduced to none or mild. All-cause mortality after 30 days was 0%.
CONCLUSIONS: It was concluded that CoreValve Evolut R is well suited for ViV implantation.
METHODS: A total of 26 patients (mean age 79.4 ± 6.1 years, 17 males and 9 females) were treated for severe prosthesis stenosis (n = 9), severe regurgitation (n = 8) or severe combination of stenosis and regurgitation (n = 9). All patients underwent transthoracic echocardiography before and after ViV implantation.
RESULTS: Valve-in-valve implantation of a CoreValve Evolut R was performed successfully in all pa-tients. The mean transaortic gradient for stenotic valves determined by transthoracic echocardiography was reduced significantly from 37.5 ± 15.3 mmHg in patients with prosthesis stenosis to 16.3 ± 8.2 mmHg (p < 0.001). In all cases with severe prosthesis regurgitation, regurgitation was reduced to none or mild. All-cause mortality after 30 days was 0%.
CONCLUSIONS: It was concluded that CoreValve Evolut R is well suited for ViV implantation.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app