Predictive Capability of an iPad-Based Medical Device (me d x) for the Diagnosis of Vertigo and Dizziness.

BACKGROUND: Making the correct diagnosis of patients presenting with vertigo and dizziness in clinical practice is often challenging.

OBJECTIVE: In this study we examined the performance of the iPad based program me d x in the prediction of different clinical vertigo and dizziness diagnoses and as a diagnostic tool to distinguish between them.

PATIENTS AND METHODS: The data collection was done in the outpatient clinic of the German Center of Vertigo and Balance Disorders. The "gold standard diagnosis" was defined as the clinical diagnosis of the specialist during the visit of the patient based on standardized history and clinical examination. Another independent and blinded physician finalized each patient's case in the constellatory diagnostic system of me d x based on an algorithm using all available clinical information. These diagnoses were compared to the "gold standard" by retrospective review of the charts of the patients. The accuracy provided by me d x was defined as the number of correctly classified diagnoses. In addition, the probability of being test positive when a disease was present (sensitivity), of being test negative when a disease was absent (specificity), of having the disease when the test is positive (positive predictive value) and of not having the disease when the test is negative (negative predictive value) for the most common diagnoses were reported. Sixteen possible different vertigo and dizziness diagnoses could be provided by me d x.

RESULTS: A total of 610 patients (mean age 58.1 ± 16.3 years, 51.2% female) were included. The accuracy for the most common diagnoses was between 82.1 and 96.6% with a sensitivity of 40 to 80.5% and a specificity of more than 80%. When analyzing the quality of me d x in a multiclass problem for the six most common clinical diagnoses, the sensitivity, specificity, positive and negative predictive values were as follows: Bilateral vestibulopathy (81.6, 97.1, 71.1, and 97.5%), Menière's disease (77.8, 97.6, 87.0, and 95.3%), benign paroxysmal positional vertigo (61.7, 98.3, 86.6, and 93.4%), downbeat nystagmus syndrome (69.6, 97.7, 71.1, and 97.5%), vestibular migraine (34.7, 97.8, 76.1, and 88.3%), and phobic postural vertigo (80.5, 82.5, 52.5, and 94.6%).

CONCLUSION: This study demonstrates that me d x is a new and easy approach to screen for different diagnoses. With the high specificity and negative predictive value, the system helps to rule out differential diagnoses and can therefore also lead to a cost reduction in the health care system. However, the sensitivity was unexpectedly low, especially for vestibular migraine. All in all, this device can only be a complementary tool, in particular for non-experts in the field.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app