Standardized model of porcine resuscitation using a custom-made resuscitation board results in optimal hemodynamic management.

AIM: Standardized modeling of cardiac arrest and cardiopulmonary resuscitation (CPR) is crucial to evaluate new treatment options. Experimental porcine models are ideal, closely mimicking human-like physiology. However, anteroposterior chest diameter differs significantly, being larger in pigs and thus poses a challenge to achieve adequate perfusion pressures and consequently hemodynamics during CPR, which are commonly achieved during human resuscitation. The aim was to prove that standardized resuscitation is feasible and renders adequate hemodynamics and perfusion in pigs, using a specifically designed resuscitation board for a pneumatic chest compression device.

METHODS AND RESULTS: A "porcine-fit" resuscitation board was designed for our experiments to optimally use a pneumatic compression device (LUCAS® II, Physio-Control Inc.), which is widely employed in emergency medicine and ideal in an experimental setting due to its high standardization. Asphyxial cardiac arrest was induced in 10 German hybrid landrace pigs and cardiopulmonary resuscitation was performed according to ERC/AHA 2015 guidelines with mechanical chest compressions. Hemodynamics were measured in the carotid and pulmonary artery. Furthermore, arterial blood gas was drawn to assess oxygenation and tissue perfusion. The custom-designed resuscitation board in combination with the LUCAS® device demonstrated highly sufficient performance regarding hemodynamics during CPR (mean arterial blood pressure, MAP 46 ± 1 mmHg and mean pulmonary artery pressure, mPAP of 36 ± 1 mmHg over the course of CPR). MAP returned to baseline values at 2 h after ROSC (80 ± 4 mmHg), requiring moderate doses of vasopressors. Furthermore, stroke volume and contractility were analyzed using pulse contour analysis (106 ± 3 ml and 1097 ± 22 mmHg/s during CPR). Blood gas analysis revealed CPR-typical changes, normalizing in the due course. Thermodilution parameters did not show persistent intravascular volume shift.

CONCLUSION: Standardized cardiopulmonary resuscitation is feasible in a porcine model, achieving adequate hemodynamics and consecutive tissue perfusion of consistent quality.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app