Bilateral branch pulmonary artery valve implantation in repaired tetralogy of fallot.

BACKGROUND: Transcatheter, bilateral branch pulmonary artery (PA) valve implantation is a novel treatment for patients with severe pulmonary insufficiency and oversized right ventricle (RV) outflow tract. There is scarce data on efficacy and safety of this approach.

METHODS: This was a retrospective study of 8 patients with repaired tetralogy of fallot (TOF) who underwent bilateral branch PA valve implantation. Demographics, echocardiography, cardiac catheterization, and axial imaging data were reviewed. Variables were compared by a paired sample t-test.

RESULTS: All patients were adult sized (weight 43-99 kg) with oversized RV outflow tract not suitable for conventional transcatheter pulmonary valve implantation. Staged bare metal PA stenting followed by valve implantation (interval 3-5 months) was technically successful in 7 patients with one stent embolization. In another patient, proximal stent migration prevented placement of bilateral pulmonary valve stents. There were a total of 14 valved branch PA stents placed (Melody valve n = 9, Sapien XT n = 2, Sapien 3 n = 3). In the 7 patients undergoing successful branch pulmonary valve placement, at median follow up of 10 months (range 3 months to 6 years), 13 (93%) valves had none/trivial insufficiency on echocardiography. Prevalve and postvalve implantation cardiac magnetic resonance imaging in five patients showed significant reduction of indexed RV end-diastolic volume (152 ± 27 to 105 ± 15 mL/m2 , P < .001).

CONCLUSIONS: Transcatheter, bilateral branch PA valve implantation was technically feasible with satisfactory efficacy and safety in patients with repaired TOF, severe pulmonary insufficiency, and oversized RV outflow tracts. Elimination of pulmonary insufficiency with this method resulted in reduced RV end-diastolic volume. This approach can be offered as an alternative to surgery, particularly in patients considered high risk for standard surgical placement and who are not candidates for the newer self-expanding valve prosthesis for placement in RV outflow tracts larger than 30 mm diameter.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app