Lack of Benefit of Heliox During Mechanical Ventilation of Subjects With Severe Air-Flow Obstruction.
Respiratory Care 2018 April
BACKGROUND: The value of heliox (helium-oxygen mixture) for patients with severe air-flow obstruction is uncertain. The purpose of this study was to determine whether heliox could reduce the degree of hyperinflation and hypercapnia in mechanically ventilated patients with severe air-flow obstruction.
METHODS: This was a single-center, prospective observational study conducted in a medical ICU of an academic medical center. We assessed the impact of heliox (65-70% helium, 30-35% oxygen) on airway pressures and arterial blood gases of 13 subjects undergoing mechanical ventilation for severe asthma ( n = 8) or exacerbation of COPD ( n = 5).
RESULTS: As compared with ventilation with air-O2 , heliox resulted in a reduction in peak airway pressure (54.1 ± 12.6 cm H2 O vs 47.9 ± 10.8 cm H2 O, P < .001) and PaCO2 (64.3 ± 14.9 mm Hg vs 62.3 + 15.1 mm Hg, P = .01). In contrast, there was no change in plateau pressure (25.3 ± 5.5 cm H2 O vs 25.8 ± 5.6 cm H2 O, P = .14) or total PEEP (13.4 ± 3.8 cm H2 O vs 13.3 ± 4.1 cm H2 O, P = .79) in response to heliox.
CONCLUSIONS: In mechanically ventilated subjects with severe air-flow obstruction, administration of heliox had no effect on indices of dynamic hyperinflation (plateau pressure and total PEEP) and resulted in only a small reduction in PaCO2 .
METHODS: This was a single-center, prospective observational study conducted in a medical ICU of an academic medical center. We assessed the impact of heliox (65-70% helium, 30-35% oxygen) on airway pressures and arterial blood gases of 13 subjects undergoing mechanical ventilation for severe asthma ( n = 8) or exacerbation of COPD ( n = 5).
RESULTS: As compared with ventilation with air-O2 , heliox resulted in a reduction in peak airway pressure (54.1 ± 12.6 cm H2 O vs 47.9 ± 10.8 cm H2 O, P < .001) and PaCO2 (64.3 ± 14.9 mm Hg vs 62.3 + 15.1 mm Hg, P = .01). In contrast, there was no change in plateau pressure (25.3 ± 5.5 cm H2 O vs 25.8 ± 5.6 cm H2 O, P = .14) or total PEEP (13.4 ± 3.8 cm H2 O vs 13.3 ± 4.1 cm H2 O, P = .79) in response to heliox.
CONCLUSIONS: In mechanically ventilated subjects with severe air-flow obstruction, administration of heliox had no effect on indices of dynamic hyperinflation (plateau pressure and total PEEP) and resulted in only a small reduction in PaCO2 .
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app