A meta-analysis of clinical electro-oculography values.

BACKGROUND: The aim of the meta-analysis was to derive a range of mean normal clinical electrooculogram (EOG) values from a systematic review of published EOG studies that followed the guidelines of the ISCEV standard for clinical electro-oculography.

METHODS: A systematic literature review was performed using four relevant databases limited to peer-reviewed articles in English between 1967 and February 2017. Studies reporting clinical EOG or FO normal values were included when the report used a standard 30° horizontal saccade, a retinal luminance of between 100 and 250 cd m-2 , and had > 10 subjects in their normative values. The search identified 1145 articles after duplicates were removed with subsequent screening of the abstracts excluding a further 1098, resulting in 47 full-text articles that were then assessed by the author (PC) with a final nine articles meeting the inclusion criteria. An overall effect estimate using inverse variance-weighted meta-analysis was performed to estimate the mean values for the light peak/dark trough ratio (LP:DT ratio) (dilated and undilated), the time to the LP, the amplitude of the LP, dark trough (DT) and the fast oscillation (FO) peak-to-trough ratio from the included studies.

RESULTS: The mean dilated LP:DT ratio was 2.35 (95% CI 2.28-2.42); undilated LP:DT ratio was 2.37 (95% CI 2.28-2.45); LP amplitude was 835 (95% CI 631-1039) µV and the mean time to the LP being 8.2 (95% CI 7.7-8.7) min. The mean DT amplitude was 358 (95% CI 292-424) µV, and the mean FO peak-to-trough ratio was 1.13 (95% CI 1.11-1.16). The results of the LP/DT ratio are drawn from studies with a mean ± standard deviation (SD) age of 34.08 ± 12.93 years for dilated and 33.65 ± 12.28 years for undilated LP/DT ratios.

CONCLUSIONS: The meta-analysis of EOG studies has generated a reference range of normal mean values for clinicians to refer to when using the ISCEV clinical EOG. It provides a potential method to generate similar data sets from published normal values in related visual electrophysiology tests.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app