Efficient noise-tolerant estimation of heart rate variability using single-channel photoplethysmography.

BACKGROUND: The feasibility of using photoplethysmography (PPG) for estimating heart rate variability (HRV) has been the subject of many recent studies with contradicting results. Accurate measurement of cardiac cycles is more challenging in PPG than ECG due to its inherent characteristics.

METHODS: We developed a PPG-only algorithm by computing a robust set of medians of the interbeat intervals between adjacent peaks, upslopes, and troughs. Abnormal intervals are detected and excluded by applying our criteria.

RESULTS: We tested our algorithm on a large database from high-risk ICU patients containing arrhythmias and significant amounts of artifact. The average difference between PPG-based and ECG-based parameters is <1% for pNN50, <1bpm for meanHR, <1ms for SDNN, <3ms for meanNN, and <4ms for SDSD and RMSSD.

CONCLUSIONS: Our performance testing shows that the pulse rate variability (PRV) parameters are comparable to the HRV parameters from simultaneous ECG recordings.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app