Spine Navigation Based on 3-Dimensional Robotic Fluoroscopy for Accurate Percutaneous Pedicle Screw Placement: A Prospective Study of 66 Consecutive Cases.

World Neurosurgery 2017 December
BACKGROUND: Minimally invasive spine surgery is associated with obstructed visibility of anatomic landmarks and increased radiation exposure, leading to higher incidence of pedicle screw mispositioning. To address these drawbacks, intraoperative 3-dimensional fluoroscopy (io3DF) and navigation are being increasingly used. We aimed to present our dedicated multifunctional hybrid operating room (HyOR) setup and evaluate the accuracy and safety of io3DF image-guided spinal navigation in transforaminal lumbar interbody fusion with percutaneous pedicle screw (PPS) placement.

METHODS: The HyOR includes a fixed 3D multiaxis robotic fluoroscopy arm that moves automatically to the preprogrammed position when needed. An initial io3DF assessment is performed to collect intraoperative images, which are automatically transferred into the navigation system. These data are used to calibrate the PPSs and insert them under computer-assisted navigation. A second io3DF is performed for verifying PPS position.

RESULTS: Between January 2014 and December 2016, 66 consecutive patients (age, 58.6 ± 14.1 years) were treated for refractory lumbar degenerative pain. Seventy-three spinal levels were treated, and 276 screws were placed, with 4.2 ± 0.76 screws per patient. There was no measurable radiation to the HyOR staff, whereas the mean radiation dose per patient was 378.3 μGym2 . The overall accuracy rate of PPS placement was 99.6%. There were no significant procedure-related complications.

CONCLUSIONS: Spine navigation based on io3DF images enabled us to avoid radiation exposure to the operating room team while delivering minimal but sufficient radiation doses to our patients. This approach achieved an accuracy rate of 99.6% for PPS placement in the safe zone, without significant complications.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app