OBJECTIVE: The primary goal of this study was to compare paramedic first pass success rate between two different video laryngoscopes and direct laryngoscopy (DL) under simulated prehospital conditions in a cadaveric model.

METHODS: This was a non-randomized, group-controlled trial in which five non-embalmed, non-frozen cadavers were intubated under prehospital spinal immobilization conditions using DL and with both the GlideScope Ranger (GL; Verathon Inc, Bothell, Washington USA) and the VividTrac VT-A100 (VT; Vivid Medical, Palo Alto, California USA). Participants had to intubate each cadaver with each of the three devices (DL, GL, or VT) in a randomly assigned order. Paramedics were given 31 seconds for an intubation attempt and a maximum of three attempts per device to successfully intubate each cadaver. Confirmation of successful endotracheal intubation (ETI) was confirmed by one of the six on-site physicians.

RESULTS: Successful ETI within three attempts across all devices occurred 99.5% of the time overall and individually 98.5% of the time for VT, 100.0% of the time for GL, and 100.0% of the time for DL. First pass success overall was 64.4%. Individually, first pass success was 60.0% for VT, 68.8% for GL, and 64.5% for DL. A chi-square test revealed no statistically significant difference amongst the three devices for first pass success rates (P=.583). Average time to successful intubation was 42.2 seconds for VT, 38.0 seconds for GL, and 33.7 for seconds for DL. The average number of intubation attempts for each device were as follows: 1.48 for VT, 1.40 for GL, and 1.42 for DL.

CONCLUSION: The was no statistically significant difference in first pass or overall successful ETI rates between DL and video laryngoscopy (VL) with either the GL or VT (adult). Hodnick R , Zitek T , Galster K , Johnson S , Bledsoe B , Ebbs D . A comparison of paramedic first pass endotracheal intubation success rate of the VividTrac VT-A 100, GlideScope Ranger, and direct laryngoscopy under simulated prehospital cervical spinal immobilization conditions in a cadaveric model. Prehosp Disaster Med. 2017;32(6):621-624.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app