Mitral annuloplasty ring suture forces: Impact of surgeon, ring, and use conditions.
Journal of Thoracic and Cardiovascular Surgery 2018 January
OBJECTIVE: The study objective was to quantify the effect of ring type, ring-annulus sizing, suture position, and surgeon on the forces required to tie down and constrain a mitral annuloplasty ring to a beating heart.
METHODS: Physio (Edwards Lifesciences, Irvine, Calif) or Profile 3D (Medtronic, Dublin, Ireland) annuloplasty rings were instrumented with suture force transducers and implanted in ovine subjects (N = 23). Tie-down forces and cyclic contractile forces were recorded and analyzed at 10 suture positions and at 3 levels of increasing peak left ventricular pressure.
RESULTS: Across all conditions, tie-down force was 2.7 ± 1.4 N and cyclic contractile force was 2.0 ± 1.2 N. Tie-down force was not meaningfully affected by any factor except surgeon. Significant differences in overall and individual tie-down forces were observed between the 2 primary implanting surgeons. No other factors were observed to significantly affect tie-down force. Contractile suture forces were significantly reduced by ring-annulus true sizing. This was driven almost exclusively by Physio cases and by reduction along the anterior aspect, where dehiscence is less common clinically. Contractile suture forces did not differ significantly between ring types. However, when undersizing, Profile 3D forces were significantly more uniform around the annular circumference. A suture's tie-down force did not correlate to its eventual contractile force.
CONCLUSIONS: Mitral annuloplasty suture loading is influenced by ring type, ring-annulus sizing, suture position, and surgeon, suggesting that reports of dehiscence may not be merely a series of isolated errors. When compared with forces known to cause suture dehiscence, these in vivo suture loading data aid in establishing potential targets for reducing the occurrence of ring dehiscence.
METHODS: Physio (Edwards Lifesciences, Irvine, Calif) or Profile 3D (Medtronic, Dublin, Ireland) annuloplasty rings were instrumented with suture force transducers and implanted in ovine subjects (N = 23). Tie-down forces and cyclic contractile forces were recorded and analyzed at 10 suture positions and at 3 levels of increasing peak left ventricular pressure.
RESULTS: Across all conditions, tie-down force was 2.7 ± 1.4 N and cyclic contractile force was 2.0 ± 1.2 N. Tie-down force was not meaningfully affected by any factor except surgeon. Significant differences in overall and individual tie-down forces were observed between the 2 primary implanting surgeons. No other factors were observed to significantly affect tie-down force. Contractile suture forces were significantly reduced by ring-annulus true sizing. This was driven almost exclusively by Physio cases and by reduction along the anterior aspect, where dehiscence is less common clinically. Contractile suture forces did not differ significantly between ring types. However, when undersizing, Profile 3D forces were significantly more uniform around the annular circumference. A suture's tie-down force did not correlate to its eventual contractile force.
CONCLUSIONS: Mitral annuloplasty suture loading is influenced by ring type, ring-annulus sizing, suture position, and surgeon, suggesting that reports of dehiscence may not be merely a series of isolated errors. When compared with forces known to cause suture dehiscence, these in vivo suture loading data aid in establishing potential targets for reducing the occurrence of ring dehiscence.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app