Ultrasound measures of tendon thickness: Intra-rater, Inter-rater and Inter-machine reliability

María Elena Del Baño-Aledo, Jacinto Javier Martínez-Payá, José Ríos-Díaz, Silvia Mejías-Suárez, Sergio Serrano-Carmona, Ana de Groot-Ferrando
Muscles, Ligaments and Tendons Journal 2017, 7 (1): 192-199
28717629

BACKGROUND: Ultrasound imaging is often used by physiotherapists and other healthcare professionals but the reliability of image acquisition with different ultrasound machines is unknown. The objective was to compare the intra-rater, inter-rater and intermachine reliability of thickness measurements of the plantar fascia (PF), Achilles tendon (AT), patellar tendon (PT) and elbow common extensor tendon (ECET) with musculoskeletal ultrasound imaging (MSUS).

METHODS: Tendon thickness was measured in four anatomical structures (14 participants, 28 images per tendon) by two sonographers and with two different ultrasound machines. Intraclass Correlation Coefficients (ICCs) and Bland-Altman plots were calculated. The standard error of measurement (SEM) and minimum detectable difference (MDD) were calculated.

RESULTS: Inter-rater reliability was excellent for AT (ICC=0.98; 95% CI= 0.96-0.99) and very good for PT (ICC=0.85; 95% CI = 0.67-0.93) and ECET (ICC=0.81; 95% CI= 0.72-0.94). Reliability for PF was moderate, with an ICC of 0.63 (CI 95%= 0.20-0.83). Bland-Altman plot for inter-machine reliability showed a mean difference of 1 m for PF measurements and a mean difference of 4 m and 20 m for AT and PT. The relative SEMs were below 7% and the MDCs were below 0.7 mm.

CONCLUSION: The MSUS reliability in measuring thickness of the four tendons is confirmed by the homogeneous readings intra sonographers, between operators and between different machines. Level of evidence: Tendon thickness can be measured reliably on different ultrasound devices, which is an important step forward in the use of this technique in daily clinical practice and research.

LEVEL OF EVIDENCE: III.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
28717629
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"